Acute and delayed outcomes of mechanical thrombectomy with use of the steerable Amplatz thrombectomy device in a model of subacute inferior vena cava thrombosis

被引:29
作者
Gu, XP
Sharafuddin, MJA
Titus, JL
Urness, M
CerveraCeballos, JJ
Ruth, GD
Amplatz, K
机构
[1] WASHINGTON UNIV, SCH MED, MALLINCKRODT INST RADIOL, SECT VASC & INTERVENT RADIOL, ST LOUIS, MO 63110 USA
[2] UNIV MINNESOTA HOSP & CLIN, DEPT RADIOL, DIV CARDIOVASC & INTERVENT RADIOL, MINNEAPOLIS, MN 55455 USA
[3] JESSE EDWARDS REGISTRY CARDIOVASC DIS, ST PAUL, MN USA
关键词
embolism; pulmonary; thrombectomy; thrombosis; venous; venae cavae;
D O I
10.1016/S1051-0443(97)70692-4
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To study the efficacy and delayed outcome of mechanical thrombectomy with the Amplatz thrombectomy device (ATD) in an experimental model of subacute inferior vena cava (IVC) thrombosis. MATERIALS AND METHODS: Mechanical thrombectomy was performed in 23 dogs with subacute infrarenal IVC thrombosis (6-15 days old), Heparin was administered during thrombectomy in all procedures (activated clotting time greater than or equal to 300 sec), Thirteen animals were killed immediately after thrombectomy, and the remaining 10 were allowed to survive for up to 1 month with no anticoagulation therapy. RESULTS: Venographic patency of the IVC was restored in all animals, although residual mural thrombus remained in nine dogs (< 20% narrowing in seven, 20%-30% narrowing in two), No histopathologic evidence of mechanical wall disruption attributed to mechanical thrombectomy was seen, However, foci of organizing residual thrombus with associated transmural phlebitic changes with round-cellular infiltration were present in all acute specimens, including those appearing clear at venography, Venography at 1 week or 1 month after thrombectomy showed IVC rethrombosis in eight dogs (80%) who were not receiving anticoagulants, During mechanical thrombectomy, a small increase in mean pulmonary artery pressure occurred, with a corresponding decrease in systemic arterial oxygen saturation, No acute emboli were noted on the post-thrombectomy pulmonary angiograms, However, histopathologic examination of acutely explanted lungs in 11 animals showed arteriolar microemboli (100-500 mu m) in four. CONCLUSION: Mechanical thrombectomy with use of the ATD can effectively clear subacute IVC thrombus, However, rethrombosis is common and may be due to the high prevalence of phlebitis and residual thrombus, Anticoagulation may need to be continued after successful thrombectomy to prevent progression of residual thrombus and allow mural phlebitic changes to subside.
引用
收藏
页码:947 / 956
页数:10
相关论文
共 44 条
[31]   MECHANICAL DISRUPTION OF PULMONARY EMBOLI IN DOGS WITH A FLEXIBLE ROTATING-TIP CATHETER (KENSEY CATHETER) [J].
STEIN, PD ;
SABBAH, HN ;
BASHA, MA ;
POPOVICH, J ;
KENSEY, KR ;
NASH, JE .
CHEST, 1990, 98 (04) :994-998
[32]   THROMBECTOMY, LYSIS, OR HEPARIN TREATMENT - CONCURRENT THERAPIES OF DEEP-VEIN THROMBOSIS - THERAPY AND EXPERIMENTAL STUDIES [J].
STIEGLER, H ;
ARBOGAST, H ;
NEES, S ;
HALDER, A ;
GRAU, A ;
RIESS, H .
SEMINARS IN THROMBOSIS AND HEMOSTASIS, 1989, 15 (03) :250-258
[33]   MECHANICAL THROMBECTOMY WITH THE AMPLATZ DEVICE - HUMAN-EXPERIENCE [J].
TADAVARTHY, SM ;
MURRAY, PD ;
INAMPUDI, S ;
NAZARIAN, GK ;
AMPLATZ, K .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1994, 5 (05) :715-724
[34]   Pulmonary emboli from pulse-spray and mechanical thrombolysis: Evaluation with an animal dialysis-graft model [J].
Trerotola, SO ;
Johnson, MS ;
Schauwecker, DS ;
Davidson, DD ;
Filo, RS ;
Zhou, XH ;
Eckert, GJ ;
Dreesen, RG ;
Carlson, KA ;
Forney, M .
RADIOLOGY, 1996, 200 (01) :169-176
[35]   Preclinical in vivo testing of a rotational mechanical thrombolytic device [J].
Trerotola, SO ;
Davidson, DD ;
Filo, RS ;
Dreesen, RG ;
Forney, M .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1996, 7 (05) :717-723
[36]   Massive pulmonary embolism: Preliminary results of treatment with the Amplatz thrombectomy device [J].
Uflacker, R ;
Strange, C ;
Vujic, I .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1996, 7 (04) :519-528
[37]   Treatment of thrombosed dialysis access grafts: Randomized trial of surgical thrombectomy versus mechanical thrombectomy with the Amplatz device [J].
Uflacker, R ;
Rajagopalan, PR ;
Vujic, I ;
Stutley, JE .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1996, 7 (02) :185-192
[38]  
Vicol C, 1996, CARDIOVASC INTER RAD, V19, P255
[39]  
Vorwerk D, 1996, CARDIOVASC INTER RAD, V19, P40
[40]   RHEOLYTIC ILIAC VENOUS THROMBECTOMY [J].
VORWERK, D ;
BUCKER, A ;
SCHMITZRODE, T ;
GUNTHER, RW .
FORTSCHRITTE AUF DEM GEBIETE DER RONTGENSTRAHLEN UND DER NEUEN BILDGEBENDEN VERFAHREN, 1995, 163 (01) :77-80