Characterization of high-sensitivity metal oxide semiconductor field effect transistor dosimeters system and LiF:Mg,Cu,P thermoluminescence dosimeters for use in diagnostic radiology

被引:33
作者
Dong, SL
Chu, TC
Lan, GY
Wu, TH
Lin, YC
Lee, JS
机构
[1] Natl Yang Ming Univ, Inst Radiol Sci, Taipei 112, Taiwan
[2] Natl Tsing Hua Univ, Dept Nucl Sci, Hsinchu, Taiwan
[3] Yuan Pei Inst Sci & Technol, Dept Radiol Technol, Taipei 112, Taiwan
关键词
diagnostic radiology; radiation detection; TLD dosimeter; MOSFET dosimeter;
D O I
10.1016/S0969-8043(02)00235-X
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
Monitoring radiation exposure during diagnostic radiographic procedures has recently become an area of interest. In recent years, the LiF:Mg,Cu,P thermoluminescence dosimeter (TLD-100H) and the highly sensitive metal oxide semiconductor field effect transistor (MOSFET) dosimeter were introduced as good candidates for entrance skin dose measurements in diagnostic radiology. In the present study, the TLD-100H and the MOSFET dosimeters were evaluated for sensitivity, linearity, energy, angular dependence, and post-exposure response. Our results indicate that the TLD-100H dosimeter has excellent linearity within diagnostic energy ranges and its sensitivity variations were under 3% at tube potentials from 40 Vp to 125 kVp. Good linearity was also observed with the MOSFET dosimeter, but in low-dose regions the values are less reliable and were found to be a function of the tube potentials. Both dosimeters also presented predictable angular dependence in this study. Our findings suggest that the TLD-100H dosimeter is more appropriate for low-dose diagnostic procedures such as chest and skull projections. The MOSFET dosimeter system is valuable for entrance skin dose measurement with lumbar spine projections and certain fluoroscopic procedures. (C) 2002 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:883 / 891
页数:9
相关论文
共 30 条
[1]  
AZORIN J, 1990, RADIAT PROT DOSIM, V33, P283
[2]   THE RESPONSE BEHAVIOR OF LIF-MG, CU, P THERMOLUMINESCENCE DOSIMETERS TO HIGH-ENERGY ELECTRON-BEAMS USED IN RADIOTHERAPY [J].
BARTOLOTTA, A ;
BRAI, M ;
CAPUTO, V ;
DILIBERTO, R ;
DIMARIANO, D ;
FERRARA, G ;
PUCCIO, P ;
SANTAMARIA, AS .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (02) :211-220
[3]   The characterization of a commercial MOSFET dosimeter system for use in diagnostic x ray [J].
Bower, MW ;
Hintenlang, DE .
HEALTH PHYSICS, 1998, 75 (02) :197-204
[4]   A new radiotherapy surface dose detector: The MOSFET [J].
Butson, MJ ;
Rozenfeld, A ;
Mathur, JN ;
Carolan, M ;
Wong, TPY ;
Metcalfe, PE .
MEDICAL PHYSICS, 1996, 23 (05) :655-658
[5]  
CHONG NS, 2000, CHINESE MED J, V63, P816
[6]   BACKSCATTER DOSE PERTURBATION IN KILOVOLTAGE PHOTON BEAMS AT HIGH ATOMIC-NUMBER INTERFACES [J].
DAS, IJ ;
CHOPRA, KL .
MEDICAL PHYSICS, 1995, 22 (06) :767-773
[7]  
DeWerd L. A., 1983, RADIAT PROT DOSIM, V6, P350
[8]   Characteristics of radiation detectors for diagnostic radiology [J].
DeWerd, LA ;
Wagner, LK .
APPLIED RADIATION AND ISOTOPES, 1999, 50 (01) :125-136
[9]  
Fleetwood DM, 1996, IEEE T NUCL SCI, V43, P779, DOI 10.1109/23.510713
[10]   AUTOMATED DATA-COLLECTION AND ANALYSIS SYSTEM FOR MOSFET RADIATION DETECTORS [J].
GLADSTONE, DJ ;
CHIN, LM .
MEDICAL PHYSICS, 1991, 18 (03) :542-548