Development of a piezoelectric polymer film sensor for plantar normal and shear stress measurements

被引:67
作者
Karki, Satu [1 ]
Lekkala, Jukka [1 ]
Kuokkanen, Hannu [2 ]
Halttunen, Jouko [1 ]
机构
[1] Tampere Univ Technol, Dept Automat Sci & Engn, FI-33101 Tampere, Finland
[2] Tampere Univ Hosp, Dept Plast Surg, FI-33521 Tampere, Finland
关键词
Piezoelectric polymer film; PVDF; Normal stress; Shear stress; Plantar pressure; PRESSURE MEASUREMENT; GAIT;
D O I
10.1016/j.sna.2009.07.010
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We have developed a new sensor prototype for plantar pressure measurements during gait. The mechanical stress at the plantar surface has two components, pressure acting normal to the surface and shear stress acting tangential to the surface. The shear stress can be further divided into anterior-posterior (AP) and medial-lateral (ML) components. The developed sensor simultaneously measures both normal and shear stresses. It utilizes commercial polyvinylidenefluoride (PVDF) material and consists of four separate sensor elements. This paper presents the sensor development and calibration for each force components. A shaker providing dynamic excitation force was used in the calibration. Average sensitivities computed from the results were (12.6 +/- 0.8)mV/N for the normal force, (223.9 +/- 20.3) mV/N for the AP shear force and (55.2 +/- 11.9) mV/N for the ML shear force. A preliminary plantar pressure measurement was also done. The results obtained were promising. However, the sensor developed here is the first prototype. In future, a matrix sensor based on this principle is planned to be constructed. (C) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:57 / 64
页数:8
相关论文
共 21 条
[11]   Simultaneous shear and pressure sensor array for assessing pressure and shear at foot/ground interface [J].
Mackey, Jeffrey R. ;
Davis, Brian L. .
JOURNAL OF BIOMECHANICS, 2006, 39 (15) :2893-2897
[12]  
Measurement Specialties Inc, 2008, PIEZ FILM SENS TECHN
[13]   Plantar pressure assessment [J].
Orlin, MN ;
McPoil, TG .
PHYSICAL THERAPY, 2000, 80 (04) :399-409
[14]  
PAAJANEN M, 2001, THESIS VTT CHEM TECH
[15]   MULTISENSOR PIEZOELECTRIC POLYMER INSOLE FOR PEDOBAROGRAPHY [J].
Pedotti, A. ;
Assente, R. ;
Fusi, G. ;
De Rossi, D. ;
Dario, P. ;
Domenici, C. .
FERROELECTRICS, 1984, 60 :163-174
[16]   Simultaneous measurement of plantar pressure and shear forces in diabetic individuals [J].
Perry, JE ;
Hall, JO ;
Davis, BL .
GAIT & POSTURE, 2002, 15 (01) :101-107
[17]  
Rahman MA., 2006, Itbm-Rbm, V27, P46, DOI [DOI 10.1016/J.RBMRET.2006.03.001, 10.1016/j.rbmret.2006.03.001]
[18]   Design, development, and characteristics of an in-shoe triaxial pressure measurement transducer utilizing a single element of piezoelectric copolymer film [J].
Razian, MA ;
Pepper, MG .
IEEE TRANSACTIONS ON NEURAL SYSTEMS AND REHABILITATION ENGINEERING, 2003, 11 (03) :288-293
[19]  
Rosenbaum D., 1997, FOOT ANKLE SURG, V3, P1, DOI DOI 10.1046/J.1460-9584.1997.00043.X
[20]   Footprint analysis of gait using a pressure sensor system [J].
Titianova, EB ;
Mateev, PS ;
Tarkka, IA .
JOURNAL OF ELECTROMYOGRAPHY AND KINESIOLOGY, 2004, 14 (02) :275-281