Reduction of eddy-current-induced distortion in diffusion MRI using a twice-refocused spin echo

被引:1037
作者
Reese, TG
Heid, O
Weisskoff, RM
Wedeen, VJ
机构
[1] Massachusetts Gen Hosp, Dept Radiol, Boston, MA 02114 USA
[2] Siemens AG, Med Engn Div, D-8520 Erlangen, Germany
[3] Epix Med Inc, Cambridge, MA USA
关键词
diffusion MRI; eddy currents; image distortion; pulse sequence; artifact;
D O I
10.1002/mrm.10308
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Image distortion due to field gradient eddy currents can create image artifacts in diffusion-weighted MR images. These images, acquired by measuring the attenuation of NMR signal due to directionally dependent diffusion, have recently been shown to be useful in the diagnosis and assessment of acute stroke and in mapping of tissue structure. This work presents an improvement on the spin-echo (SE) diffusion sequence that displays less distortion and consequently improves image quality. Adding a second refocusing pulse provides better image quality with less distortion at no cost in scanning efficiency or effectiveness, and allows more flexible diffusion gradient timing. By adjusting the timing of the diffusion gradients, eddy currents with a single exponential decay constant can be nulled, and eddy currents with similar decay constants can be greatly reduced. This new sequence is demonstrated in phantom measurements and in diffusion anisotropy images of normal human brain. (C) 2003 Wiley-Liss, Inc.