3-Nitrotyrosine, a product of tyrosine nitration, is useful as a marker for the generation of reactive nitrogen oxide species with short half-lives such as peroxynitrite. A reverse-phase high-pressure liquid chromatographic method using a dual-mode electrochemical detector in series with a photodiode array detector has been developed to determine the levels of 3-nitrotyrosine in biological samples. The principle of this method involves reduction of 3-nitrotyrosine at an upstream gold amalgam electrode and oxidation of the resulting product(s) at a downstream glassy carbon electrode. 3-Nitrotyrosine is quantified by the amount of the current generated at the downstream electrode, and a femtomole detection level can be achieved. The disappearance of the corresponding peak when the electrochemical detector is used only in the single oxidative:mode provides additional evidence for the identity of 3-nitrotyrosine in the sample. Tyrosine from the same sample is determined by its UV absorption at 280 nm, thus eliminating the need for an internal standard. With this method a dose-dependent increase of 3- to 10-fold in the levels of protein 3-nitrotyrosine was observed in the blood plasma, and a 2- to 4-fold increase in the lung cytosols, of rats treated with the lung carcinogen and nitrating agent tetranitromethane. (C) 2000 Academic Press.