High-performance fiber optic pH microsensors for practical physiological measurements using a dual-emission sensitive dye

被引:145
作者
Song, A [1 ]
Parus, S [1 ]
Kopelman, R [1 ]
机构
[1] UNIV MICHIGAN, DEPT CHEM, ANN ARBOR, MI 48109 USA
关键词
D O I
10.1021/ac960917+
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A fast and durable ratiometric pH microoptode that is highly accurate, precise, sensitive, reversible, and reproducible over the physiological ranges of pH, ionic strength, and temperature has been developed. The sensing site consists of 5 (and 6)-carboxynaphthofluorescein (CNF) entrapped in a polyacrylamide gel matrix via photopolymerization at the silanized end of an optical fiber with a diameter of 2 (pulled) or 125 mu m (unpulled). The optode's precision for the pH 6.3-8.4 range in rat embryos, sera, or physiological (Earle's and Tyrode's) buffers was found to be better than +/-0.03 pH unit. The pulled and unpulled optodes have respective upper limit response times of 1 and 400 ms for 1-pH-unit change. Over a 7-week period, they retain sensitivity for 600 and 10 000 measurements, respectively. Ratiometric measurements are made using a ps-sensitive emission peak on each side of an isosbestic point. The CNF microoptode is most suitable for biological applications because of its essentially linear response over the pH 7-8 range, its high sensitivity (slope about 2), and its almost perfect correlation with a pH macroelectrode. Furthermore, errors introduced by photobleaching, leaching, quenching, optode movement, and excitation source fluctuations are minimal.
引用
收藏
页码:863 / 867
页数:5
相关论文
共 38 条
  • [1] MODULATION OF PH BY NEURONAL-ACTIVITY
    CHESLER, M
    KAILA, K
    [J]. TRENDS IN NEUROSCIENCES, 1992, 15 (10) : 396 - 402
  • [2] Clark C H, 1992, Am J Crit Care, V1, P53
  • [3] DECREASING PH OF RAT EMBRYOS AND FLUIDS ESTIMATED BY TRANS-PLACENTAL DISTRIBUTION OF DMO
    COLLINS, MD
    DUGGAN, CA
    SCHREINER, CM
    SCOTT, WJ
    [J]. AMERICAN JOURNAL OF PHYSIOLOGY, 1989, 257 (03): : R542 - R549
  • [4] APPLICATION OF A NOVEL FIBEROPTIC BIOSENSOR IN-SITU TO INVESTIGATE THE METABOLIC EFFECT OF LACTATE INFUSION
    DAGER, SR
    YIM, JB
    KHALIL, GE
    ARTRU, AA
    BOWDEN, DM
    KENNY, MA
    [J]. NEUROPSYCHOPHARMACOLOGY, 1995, 12 (04) : 307 - 313
  • [5] DOURADO S, IN PRESS P SPIE INT
  • [6] FANG ZL, 1984, ANAL CHIM ACTA, V164, P41
  • [7] PEAK VELOCITY OF THE OUTFLOW TRACT OF THE AORTA - CORRELATIONS WITH ACID-BASE STATUS AND OXYGENATION OF THE GROWTH-RETARDED FETUS
    FERRAZZI, E
    BELLOTTI, M
    MARCONI, A
    FLISI, L
    BARBERA, A
    PARDI, G
    [J]. OBSTETRICS AND GYNECOLOGY, 1995, 85 (05) : 663 - 668
  • [8] GASTRIC INTRAMUCOSAL PH, TISSUE OXYGENATION AND ACID-BASE-BALANCE
    FIDDIANGREEN, RG
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 1995, 74 (05) : 591 - 606
  • [9] GEKLE M, 1995, J PHARMACOL EXP THER, V275, P397
  • [10] Gow Neil A. R., 1993, Trends in Microbiology, V1, P45, DOI 10.1016/0966-842X(93)90031-L