A simplified model of the source channel of the Leksell GammaKnife® tested with PENELOPE

被引:25
作者
Al-Dweri, FM [1 ]
Lallena, AM
Vilches, M
机构
[1] Univ Granada, Dept Fis Moderna, E-18071 Granada, Spain
[2] Hosp Clin San Cecilio, Serv Radiofis, E-18012 Granada, Spain
关键词
D O I
10.1088/0031-9155/49/12/015
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Monte Carlo simulations using the code PENELOPE have been performed to test a simplified model of the source channel geometry of the Leksell GammaKnife(R). The characteristics of the radiation passing through the treatment helmets are analysed in detail. We have found that only primary particles emitted from the source with polar angles smaller than 3 with respect to the beam axis are relevant for the dosimetry of the Gamma Knife. The photon trajectories reaching the output helmet collimators at (x, y, z = 236 mm) show strong correlations between rho = (x(2) + y(2))(1/2) and their polar angle theta, on one side, and between tan(-1)(y/x) and their azimuthal angle phi, on the other. This enables us to propose a simplified model which treats the full source channel as a mathematical collimator. This simplified model produces doses in good agreement with those found for the full geometry. In the region of maximal dose, the relative differences between both calculations are within 3%, for the 18 and 14 mm helmets, and 10%, for the 8 and 4 mm ones. Besides, the simplified model permits a strong reduction (larger than a factor 15) in the computational time.
引用
收藏
页码:2687 / 2703
页数:17
相关论文
共 16 条
[1]  
BENJAMIN WC, 1997, SEMIN ONCOL, V24, P707
[2]   Monte Carlo calculation of single-beam dose profiles used in a gamma knife treatment planning system [J].
Cheung, JYC ;
Yu, KN ;
Yu, CP ;
Ho, RTK .
MEDICAL PHYSICS, 1998, 25 (09) :1673-1675
[3]   Monte Carlo calculations and GafChromic film measurements for plugged collimator helmets of Leksell Gamma Knife unit [J].
Cheung, JYC ;
Yu, KN ;
Ho, RTK ;
Yu, CP .
MEDICAL PHYSICS, 1999, 26 (07) :1252-1256
[4]   Monte Carlo calculated output factors of a Leksell Gamma Knife unit [J].
Cheung, JYC ;
Yu, KN ;
Ho, RTK ;
Yu, CP .
PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (12) :N247-N249
[5]   Dose distributions at extreme irradiation depths of gamma knife radiosurgery: EGS4 Monte Carlo calculations [J].
Cheung, JYC ;
Yu, KN ;
Yu, CP ;
Ho, RTK .
APPLIED RADIATION AND ISOTOPES, 2001, 54 (03) :461-465
[6]   Stereotactic dose planning system used in Leksell Gamma Knife model-B: EGS4 Monte Carlo versus GafChromic films MD-55 [J].
Cheung, YC ;
Yu, KN ;
Ho, RTK ;
Yu, CP .
APPLIED RADIATION AND ISOTOPES, 2000, 53 (03) :427-430
[7]  
*EL, 1996, LEKS GAMM PLAN INSTR
[8]  
*EL INSTR AB, 1992, LEKS GAMM UN US MAN
[9]  
Hartmann G, 1995, QUALITY ASSURANCE PR
[10]   Simulation of dose distribution irradiation by the Leksell Gamma Unit [J].
Liu, XW ;
Zhang, CX .
PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (02) :441-445