A method to create an optimum current distribution and homogeneous B-1 field for elliptical birdcage coils

被引:22
作者
Li, SZ
Collins, CM
Dardzinski, BJ
Chin, CL
Smith, MB
机构
[1] PENN STATE UNIV,COLL MED,DEPT RADIOL,CTR NMR RES,HERSHEY,PA 17033
[2] PENN STATE UNIV,COLL MED,DEPT CELLULAR & MOL PHYSIOL,HERSHEY,PA 17033
[3] UNIV PENN,DEPT BIOENGN,PHILADELPHIA,PA 19104
[4] PENN STATE UNIV,BIOENGN PROGRAM,UNIVERSITY PK,PA 16802
关键词
birdcage coil; B-1 field homogeneity;
D O I
10.1002/mrm.1910370420
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
For a birdcage coil with elliptical cross-section, a sinusoidal current pattern does not provide a homogeneous B-1 field, A simple theory was developed to create an optimized current distribution for elliptical birdcage coils, This optimized current pattern can create a perfectly homogeneous B-1 field inside any elliptical shape, To verify the theory, a 16-element high-pass elliptical birdcage coil was built inside a circular RF shield, The current was optimized by using the inductance characteristics of the coil components to calculate the end-ring capacitances. The B-1 field was theoretically calculated and experimentally mapped for the optimized elliptical birdcage coil and a nonoptimized coil, The results demonstrate that by optimizing the current distribution, a very homogeneous B-1 field is produced, This method can be directly applied in design and construction of elliptical birdcage coils for imaging of the naturally occurring elliptical cross-sectional geometries in the human body.
引用
收藏
页码:600 / 608
页数:9
相关论文
共 14 条
[1]  
BECK BL, 1994, P SMR 2 ANN M SAN FR, P1111
[2]  
Bimson W.E., 1992, P INT SOC MAGNETIC R, P272
[3]  
GROVER RW, 1946, INDUCTANCE CALCULATI
[4]   AN EFFICIENT, HIGHLY HOMOGENEOUS RADIOFREQUENCY COIL FOR WHOLE-BODY NMR IMAGING AT 1.5-T [J].
HAYES, CE ;
EDELSTEIN, WA ;
SCHENCK, JF ;
MUELLER, OM ;
EASH, M .
JOURNAL OF MAGNETIC RESONANCE, 1985, 63 (03) :622-628
[5]  
HAYES CE, 1989, P SMRM 8 ANN M AMST, P974
[6]   A TECHNIQUE FOR DOUBLE RESONANT OPERATION OF BIRDCAGE IMAGING COILS [J].
JOSEPH, PM ;
LU, DF .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1989, 8 (03) :286-294
[7]  
KURCZEWSKI R, 1992, P SMRM 11 ANN M BERL, P4025
[8]  
LI CS, 1993, P SMRM 12 ANN M NEW, P1342
[9]   RF COIL OPTIMIZATION - EVALUATION OF B-1 FIELD HOMOGENEITY USING FIELD HISTOGRAMS AND FINITE-ELEMENT CALCULATIONS [J].
LI, SZ ;
YANG, QX ;
SMITH, MB .
MAGNETIC RESONANCE IMAGING, 1994, 12 (07) :1079-1087
[10]  
LU N, 1995, P SMR 3 ANN M NIC, P1002