Flat-panel digital radiology with amorphous selenium and active-matrix readout

被引:38
作者
Rowlands, JA [1 ]
Zhao, W [1 ]
Blevis, IM [1 ]
Waechter, DF [1 ]
Huang, ZS [1 ]
机构
[1] LITTON SYST CANADA, ETOBICOKE, ON, CANADA
关键词
radiography; digital; technology; screens and films; selenium;
D O I
10.1148/radiographics.17.3.9153709
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Future imaging techniques will be capable of high image quality, fast acquisition, compactness, and versatile operation, A flat-panel imager flat is expected to achieve these goals is under development, It consists of a thin layer of amorphous selenium that converts xrays directly to an electric charge and a thin-film electronic circuit, or active matrix, to read out the electronic signal directly to a computer host. The advantages of amorphous selenium include high resolution and low noise without loss of signal strength. The advantages of the active matrix are real-time readout, flexible design parameters, and compactness of the readout structures, A prototype of this imager has been built and operated, and initial images (of an x-ray lest bar pattern and a hand phantom) have been acquired. Although the prototype was built to test scientific principles and many possibilities for optimization remain, the images already possess the quality necessary for many radiographic procedures. A large range of current and new applications exist for this imager.
引用
收藏
页码:753 / 760
页数:8
相关论文
共 16 条
[1]   A REAL-TIME, FLAT-PANEL, AMORPHOUS-SILICON, DIGITAL X-RAY IMAGER [J].
ANTONUK, LE ;
YORKSTON, J ;
HUANG, WD ;
SIEWERDSEN, JH ;
BOUDRY, JM ;
ELMOHRI, Y ;
MARX, MV .
RADIOGRAPHICS, 1995, 15 (04) :993-1000
[2]   XERORADIOGRAPHY [J].
BOAG, JW .
PHYSICS IN MEDICINE AND BIOLOGY, 1973, 18 (01) :3-37
[3]   RADIOGRAPHIC INFORMATION-THEORY AND APPLICATION TO MAMMOGRAPHY [J].
BRODIE, I ;
GUTCHECK, RA .
MEDICAL PHYSICS, 1982, 9 (01) :79-95
[4]   Amorphous silicon x-ray image sensor [J].
Chabbal, J ;
Chaussat, C ;
Ducourant, T ;
Fritsch, L ;
Michailos, J ;
Spinnler, V ;
Vieux, G ;
Arques, M ;
Hahm, G ;
Hoheisel, M ;
Horbaschek, H ;
Schulz, R ;
Spahn, M .
PHYSICS OF MEDICAL IMAGING: MEDICAL IMAGING 1996, 1996, 2708 :499-510
[5]   IMAGE FEATURE ANALYSIS AND COMPUTER-AIDED DIAGNOSIS IN DIGITAL RADIOGRAPHY .1. AUTOMATED DETECTION OF MICROCALCIFICATIONS IN MAMMOGRAPHY [J].
CHAN, HP ;
DOI, K ;
GALHOTRA, S ;
VYBORNY, CJ ;
MACMAHON, H ;
JOKICH, PM .
MEDICAL PHYSICS, 1987, 14 (04) :538-548
[6]   High-resolution amorphous silicon image sensor [J].
Graeve, T ;
Li, YM ;
Fabans, A ;
Huang, WG .
PHYSICS OF MEDICAL IMAGING: MEDICAL IMAGING 1996, 1996, 2708 :494-498
[7]  
LEE DL, 1995, P SOC PHOTO-OPT INS, V2432, P237, DOI 10.1117/12.208342
[8]   IMAGE QUALITY OF A DIGITAL CHEST RADIOGRAPHY SYSTEM BASED ON A SELENIUM DETECTOR [J].
NEITZEL, U ;
MAACK, I ;
GUNTHERKOHFAHL, S .
MEDICAL PHYSICS, 1994, 21 (04) :509-516
[9]   A STUDY OF PHOTOELECTRONIC DIGITAL RADIOLOGY .3. IMAGE ACQUISITION COMPONENTS AND SYSTEM-DESIGN [J].
NUDELMAN, S ;
ROEHRIG, H ;
CAPP, MP .
PROCEEDINGS OF THE IEEE, 1982, 70 (07) :715-727
[10]  
ROUGEOT H, 1993, AAPM MONOGRAPH, V22, P49