Sentinel node mapping affects intraoperative pulse oximetric recordings during breast cancer surgery

被引:22
作者
Koivusalo, AM
Von Smitten, K
Lindgren, L
机构
[1] Helsinki Univ Hosp, Dept Med 4, Dept Anesthesia, FIN-00170 Helsinki, Finland
[2] Helsinki Univ Hosp, Dept Med 4, Dept Surg, FIN-00170 Helsinki, Finland
关键词
sentinel node; patent blue vital dye; peripheral oxygen saturation;
D O I
10.1034/j.1399-6576.2002.460414.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background: In invasive breast cancer lymphatic mapping with patent blue vital dye (PBV) is used intraoperatively to identify the sentinel lymph nodes: the first axillary node draining the mammary lymphatic basin and first involved by the metastatic growth in breast cancer. Patent blue vital dye spreads to tissues giving a bluish tinge to patients. We have noted the possibility that intraoperative peripheral pulse oximetric (SpO(2)) values are artificially low when intradermal PBV is used. Methods: Twenty patients with normal pulmonary function undergoing breast cancer surgery in standardized anesthesia either did or did not receive intradermal PBV sentinel node marking. The radial artery was cannulated for blood-gas-analysis; arterial oxygen tension (PaO2); and arterial oxygen saturation (SaO(2)). Peripheral oxygen saturation was measured using the light absorption technique. Red and infrared light (660 and 900nm), used by pulseoxymetry, is partially absorbed when passing through the tissue. The amount of light absorbed is sensored and saturation calculated. The color of the skin was evaluated. Results: Peripheral oxygen saturation decreased only immediately after the injection of PBV, and remained at a significantly lower level (P<0.001) throughout the operation and up to 90 min postoperatively. Arterial oxygen tension and SaO(2) values did not decrease after intradermal PBV Patent blue vital dye made patients' skin more bluish (P<0.001). No changes in SpO(2), PaO2 and SaO(2) were found in control patients. Conclusion: The spectrum of PBV has a peak absorption at 640 nm, thus making the SPO2 values incorrect. Peripheral oxygen saturation values are falsely low and true arterial oxygenation is not impaired when PBV is used during sentinel node mapping. (C) Acta Anaesthesiologica Scandinavica 46 (2002).
引用
收藏
页码:411 / 414
页数:4
相关论文
共 16 条
[1]   Lymphatic mapping and sentinel node biopsy in the patient with breast cancer [J].
Albertini, JJ ;
Lyman, GH ;
Cox, C ;
Yeatman, T ;
Balducci, L ;
Ku, NN ;
Shivers, S ;
Berman, C ;
Wells, K ;
Rapaport, D ;
Shons, A ;
Horton, J ;
Greenberg, H ;
Nicosia, S ;
Clark, R ;
Cantor, A ;
Reintgen, DS .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1996, 276 (22) :1818-1822
[2]   Do all patients with sentinel node metastasis from breast carcinoma need complete axillary node dissection? [J].
Chu, KU ;
Turner, RR ;
Hansen, NM ;
Brennan, MB ;
Bilchik, A ;
Giuliano, AE .
ANNALS OF SURGERY, 1999, 229 (04) :536-541
[3]   Credentialing for breast lymphatic mapping: How many cases are enough? [J].
Cody, HS ;
Hill, ADK ;
Tran, KN ;
Brennan, MF ;
Borgen, PI .
ANNALS OF SURGERY, 1999, 229 (05) :723-728
[4]   LYMPHATIC MAPPING AND SENTINEL LYMPHADENECTOMY FOR BREAST-CANCER [J].
GIULIANO, AE ;
KIRGAN, DM ;
GUENTHER, JM ;
MORTON, DL .
ANNALS OF SURGERY, 1994, 220 (03) :391-401
[5]   SPURIOUS PULSE OXIMETER DESATURATION WITH METHYLENE-BLUE INJECTION [J].
KESSLER, MR ;
EIDE, T ;
HUMAYUN, B ;
POPPERS, PJ .
ANESTHESIOLOGY, 1986, 65 (04) :435-436
[6]   THE INFLUENCE OF PATENT-BLUE-V ON PULSE OXIMETRY AND HAEMOXIMETRY [J].
LARSEN, VH ;
FREUDENDALPEDERSEN, A ;
FOGHANDERSEN, N .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1995, 39 :53-55
[7]   Sentinel lymph node biopsy in breast cancer: Unfiltered radioisotope is superior to filtered [J].
Linehan, DC ;
Hill, ADK ;
Tran, KN ;
Yeung, H ;
Yeh, SDJ ;
Borgen, PI ;
Cody, HS .
JOURNAL OF THE AMERICAN COLLEGE OF SURGEONS, 1999, 188 (04) :377-381
[8]   FALSE DESATURATION DUE TO INTRADERMAL PATENT BLUE-DYE [J].
MORELL, RC ;
HEYNEKER, T ;
KASHTAN, HI ;
RUPPE, C .
ANESTHESIOLOGY, 1993, 78 (02) :363-364
[9]  
NEWTON DW, 1981, PHARM SCI, V70, P122
[10]   POTENTIAL ERRORS IN PULSE OXIMETRY .3. EFFECTS OF INTERFERENCE, DYES, DYSHAEMOGLOBINS AND OTHER PIGMENTS [J].
RALSTON, AC ;
WEBB, RK ;
RUNCIMAN, WB .
ANAESTHESIA, 1991, 46 (04) :291-295