POTENTIAL ERRORS IN PULSE OXIMETRY .3. EFFECTS OF INTERFERENCE, DYES, DYSHAEMOGLOBINS AND OTHER PIGMENTS

被引:127
作者
RALSTON, AC [1 ]
WEBB, RK [1 ]
RUNCIMAN, WB [1 ]
机构
[1] ROYAL ADELAIDE HOSP,DEPT ANAESTHESIA & INTENS CARE,N TERRACE,ADELAIDE,SA 5000,AUSTRALIA
关键词
EQUIPMENT; PULSE OXIMETER; MEASUREMENT;
D O I
10.1111/j.1365-2044.1991.tb11501.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Electrosurgery, patient motion and some types of lighting can cause errors in saturation readout; it is recommended that probes should be shielded from ambient lighting. Intravenous dyes can introduce gross but transient errors, which may also be present in in vitro measurements. Carboxyhaemoglobin causes overestimation of fractional saturation by an amount less than, but possibly close to, the percent of carboxyhaemoglobin present. Methaemoglobin causes the pulse oximeter readout to tend towards 85%. Fetal haemoglobin and bilirubin introduce no significant error, although they may interfere with in vitro measurements. Skin pigmentation can result in a slight decrease in accuracy. Nail polish may cause up to 6% underestimation of saturation; it is recommended that probes should be mounted sideways on fingers with nail polish or long nails. Adhesive tape or a vinyl glove across the probe has no demonstrable effect on accuracy. A blood sample should be analysed by a multiwavelength in vitro oximeter when an erroneous pulse oximeter reading is suspected, although errors may be introduced in the in vitro reading by fetal haemoglobin, bilirubin and intravenous dyes. Copyright © 1991, Wiley Blackwell. All rights reserved
引用
收藏
页码:291 / 295
页数:5
相关论文
共 49 条
[1]   THE USE OF A VINYL GLOVE DOES NOT AFFECT PULSE OXIMETER MONITORING [J].
ACKERMAN, WE ;
JUNEJA, M ;
BAUMANN, RC ;
KACZOROWSKI, DM .
ANESTHESIOLOGY, 1989, 70 (03) :558-559
[2]  
ALEXANDER CM, 1989, ANESTH ANALG, V68, P368
[3]  
AMAZON K, 1981, AM J CLIN PATHOL, V75, P519
[4]   BENZOCAINE-INDUCED METHEMOGLOBINEMIA IN AN ADULT - ACCURACY OF PULSE OXIMETRY WITH METHEMOGLOBINEMIA [J].
ANDERSON, ST ;
HAJDUCZEK, J ;
BARKER, SJ .
ANESTHESIA AND ANALGESIA, 1988, 67 (11) :1099-1101
[5]  
[Anonymous], OPERATORS MANUAL
[6]   PULSE OXIMETRY - APPLICATIONS AND LIMITATIONS [J].
BARKER, SJ ;
TREMPER, KK .
INTERNATIONAL ANESTHESIOLOGY CLINICS, 1987, 25 (03) :155-175
[7]   THE EFFECT OF CARBON-MONOXIDE INHALATION ON PULSE OXIMETRY AND TRANS-CUTANEOUS PO2 [J].
BARKER, SJ ;
TREMPER, KK .
ANESTHESIOLOGY, 1987, 66 (05) :677-679
[8]   A CLINICAL-EVALUATION OF THE ACCURACY OF THE NELLCOR-N-100 AND OHMEDA-3700 PULSE OXIMETERS [J].
CECIL, WT ;
THORPE, KJ ;
FIBUCH, EE ;
TUOHY, GF .
JOURNAL OF CLINICAL MONITORING, 1988, 4 (01) :31-36
[9]   A COMPARISON OF THE PERFORMANCE OF 20 PULSE OXIMETERS UNDER CONDITIONS OF POOR PERFUSION [J].
CLAYTON, DG ;
WEBB, RK ;
RALSTON, AC ;
DUTHIE, D ;
RUNCIMAN, WB .
ANAESTHESIA, 1991, 46 (01) :3-10
[10]   PULSE OXIMETER PROBES - A COMPARISON BETWEEN FINGER, NOSE, EAR AND FOREHEAD PROBES UNDER CONDITIONS OF POOR PERFUSION [J].
CLAYTON, DG ;
WEBB, RK ;
RALSTON, AC ;
DUTHIE, D ;
RUNCIMAN, WB .
ANAESTHESIA, 1991, 46 (04) :260-265