1000 h stability test of the Au/Pd thermocouple at 1000 degrees C

被引:7
作者
Kim, YG
Gam, KS
Rhee, C
机构
[1] Division of Quantum Metrology, Temperature Group, Korea Res. Inst. of Std. and Science, Yusong, Taejon 305-600
关键词
D O I
10.1088/0026-1394/33/4/7
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A newly developed Au/Pd thermocouple has shown very high reproducibility and good immersion characteristics. A stability test of the Au/Pd thermocouple has been conducted for 1000 h at 1000 degrees C in air to assess its suitability for practical application. Two thermocouples from two manufacturers were fabricated and their emf values were checked in a silver freezing-point cell before and after the soaking at 1000 OC. It was found that the high-temperature stability of the Au/Pd thermocouple is within +/-30 mK (2 sigma) throughout the test time. This result implies that the Au/Pd thermocouple possesses characteristics which may make it suitable for use as a secondary high-temperature thermometer in industry.
引用
收藏
页码:337 / 339
页数:3
相关论文
共 9 条
[1]   A STUDY OF THERMOCOUPLE STABILITY, REPRODUCIBILITY AND ACCURACY (PT VS PT-RH AND PT VS AU) [J].
ANCSIN, J .
METROLOGIA, 1991, 28 (04) :339-347
[2]  
Bentley R. E., 1980, High Temperatures - High Pressures, V12, P33
[3]  
Blatt FJ, 1976, THERMOELECTRIC POWER, DOI 10.1007/978-1-4613-4268-7
[4]  
BURNS GW, 1992, TEMP MEAS C, V6, P531
[5]   A GOLD PLATINUM THERMOCOUPLE REFERENCE TABLE [J].
GOTOH, M ;
HILL, KD ;
MURDOCK, EG .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1991, 62 (11) :2778-2791
[6]  
JUNG HJ, 1982, TEMP ITS MEAS, V5, P763
[7]   IMMERSION CHARACTERISTICS AND REPRODUCIBILITY OF THE GOLD/PALLADIUM THERMOCOUPLE [J].
KIM, YG ;
GAM, KS ;
KANG, KH .
METROLOGIA, 1994, 31 (02) :145-147
[8]  
MCLAREN EH, 1987, PT AU THERMOCOUPLE
[9]  
STROUSE GF, 1992, TEMP MEAS C, V6, P389