The comparison of plasma deproteinization methods for the detection of low-molecular-weight metabolites by 1H nuclear magnetic resonance spectroscopy

被引:126
作者
Daykin, CA
Foxall, PJD
Connor, SC
Lindon, JC
Nicholson, JK
机构
[1] Univ London Imperial Coll Sci Technol & Med, Fac Med, Div Biomed Sci, London SW7 2AZ, England
[2] UCL, Sch Med, Inst Urol & Nephrol, London W1P 7PN, England
[3] GlaxoSmithKline Inc, Dept Safety Assessment, Ware SG12 0DP, Herts, England
基金
英国工程与自然科学研究理事会;
关键词
D O I
10.1006/abio.2002.5637
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Blood plasma is the major vehicle by which metabolites are transported around the body in mammalian species, and chemical analysis of plasma can provide a wealth of information relating to the biochemical status of an individual and is important for diagnostic purposes. However, plasma is very complex in physicochemical terms because it is composed of a range of organic and inorganic constituents with a wide range of molecular weights and chemical classes and this makes analysis non-trivial. It is now well established that high-resolution (1)H NMR spectroscopy of blood plasma provides useful qualitative and quantitative biochemical information relating to metabolic disorders. However, one of the problems encountered in NMR spectroscopic analysis of blood plasma is the extensive peak overlap or presence of broad macromolecule peaks in the (1)H NMR spectrum, which can severely limit the amount of obtainable information. Even with spectroscopic editing, information relating to low-molecular-weight (MW) metabolites is frequently lost. Therefore, the efficiency of a range of conventional protein removal methods, in combination with the use of one- and two-dimensional NMR spectroscopic methods for evaluation, have been compared for the extraction of NMR-observable low-MW metabolites. It has been shown that these "deproteinization" methods vary considerably in recovery of low MW metabolites and a judicious choice is crucial for optimal extraction of a given analyte. The results presented here show that while ultrafiltration provides the "safest" method of plasma deproteinization, the signal-to-noise ratio of the resultant (1)H NMR spectra is poor. On the other hand, acetonitrile precipitation at physiological pH allows the detection of more low-MW metabolites and at higher concentrations than any other method and provides the further advantages of being a rapid and simple procedure. (C) 2002 Elsevier Science (USA).
引用
收藏
页码:220 / 230
页数:11
相关论文
共 20 条
  • [1] SIMULTANEOUS DETERMINATION OF ALLANTOIN AND OXYPURINES IN BIOLOGICAL-FLUIDS BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY
    BALCELLS, J
    GUADA, JA
    PEIRO, JM
    PARKER, DS
    [J]. JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1992, 575 (01): : 153 - 157
  • [2] BALES JR, 1984, CLIN CHEM, V30, P1631
  • [3] NMR-INVISIBLE LACTATE IN BLOOD-PLASMA
    BELL, JD
    BROWN, JCC
    KUBAL, G
    SADLER, PJ
    [J]. FEBS LETTERS, 1988, 235 (1-2) : 81 - 86
  • [4] DETECTION OF ALUMINUM(III) BINDING TO CITRATE IN HUMAN BLOOD-PLASMA BY PROTON NUCLEAR-MAGNETIC-RESONANCE SPECTROSCOPY
    BELL, JD
    KUBAL, G
    RADULOVIC, S
    SADLER, PJ
    TUCKER, A
    [J]. ANALYST, 1993, 118 (03) : 241 - 244
  • [5] BOCK JL, 1982, CLIN CHEM, V28, P1873
  • [6] SPIN-ECHO PROTON NMR-SPECTROSCOPY OF URINE SAMPLES - WATER SUPPRESSION VIA A UREA-DEPENDENT T2 RELAXATION PROCESS
    CONNOR, S
    EVERETT, J
    NICHOLSON, JK
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1987, 4 (05) : 461 - 470
  • [7] GARTLAND KPR, 1988, THESIS U LONDON
  • [8] High-resolution diffusion and relaxation edited one- and two-dimensional H-1 NMR spectroscopy of biological fluids
    Liu, ML
    Nicholson, JK
    London, JC
    [J]. ANALYTICAL CHEMISTRY, 1996, 68 (19) : 3370 - 3376
  • [9] Nicholson J K, 1989, NMR Biomed, V2, P77, DOI 10.1002/nbm.1940020207
  • [10] 750-MHZ H-1 AND H-1-C-13 NMR-SPECTROSCOPY OF HUMAN BLOOD-PLASMA
    NICHOLSON, JK
    FOXALL, PJD
    SPRAUL, M
    FARRANT, RD
    LINDON, JC
    [J]. ANALYTICAL CHEMISTRY, 1995, 67 (05) : 793 - 811