Effects of turbulent stresses upon mechanical hemolysis: Experimental and computational analysis

被引:146
作者
Kameneva, MV
Burgreen, GW
Kono, K
Repko, B
Antaki, JF
Umezu, M
机构
[1] Univ Pittsburgh, McGowan Inst Regenerat Med, Pittsburgh, PA 15219 USA
[2] Univ Pittsburgh, Dept Bioengn, Pittsburgh, PA 15219 USA
[3] Optimal LLC, Starkville, MS USA
[4] Waseda Univ, Dept Mech Engn, Tokyo, Japan
[5] Carnegie Mellon Univ, Biomed Engn & Comp Sci Dept, Pittsburgh, PA 15213 USA
关键词
D O I
10.1097/01.MAT.0000136512.36370.B5
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Experimental and computational studies were performed to elucidate the role of turbulent stresses in mechanical blood damage (hemolysis). A suspension of bovine red blood cells (RBC) was driven through a closed circulating loop by a centrifugal pump. A small capillary tube (inner diameter 1 mm and length 70 mm) was incorporated into the circulating loop via tapered connectors. The suspension of RBCs was diluted with saline to achieve an asymptotic apparent viscosity of 2.0 +/- 0.1 cP at 23degreesC to produce turbulent flow at nominal flow rate and pressure. To study laminar flow at the identical wall shear stresses in the same capillary tube, the apparent viscosity of the RBC suspension was increased to 6.3 +/- 0.1 cP (at 23degreesC) by addition of Dextran-40. Using various combinations of driving pressure and Dextran mediated adjustments in dynamic viscosity Reynolds numbers ranging from 300-5,000 were generated, and rates of hemolysis were measured. Pilot studies were performed to verify that the suspension media did not affect mechanical fragility of the RBCs. The results of these bench studies demonstrated that, at the same wall shear stress in a capillary tube, the level of hemolysis was significantly greater (p < 0.05) for turbulent flow as compared with laminar flow. This confirmed that turbulent stresses contribute strongly to blood mechanical trauma. Numerical predictions of hemolysis obtained by computational fluid dynamic modeling were in good agreement with these experimental data.
引用
收藏
页码:418 / 423
页数:6
相关论文
共 39 条
[1]  
Akutsu T, 1997, ARTIF ORGANS, V21, P1110
[2]   COMPUTATIONAL FLOW OPTIMIZATION OF ROTARY BLOOD PUMP COMPONENTS [J].
ANTAKI, JF ;
GHATTAS, O ;
BURGREEN, GW ;
HE, BC .
ARTIFICIAL ORGANS, 1995, 19 (07) :608-615
[3]  
Baldwin J T, 1990, ASAIO Trans, V36, pM274
[4]   LDA MEASUREMENTS OF MEAN VELOCITY AND REYNOLDS STRESS-FIELDS WITHIN AN ARTIFICIAL-HEART VENTRICLE [J].
BALDWIN, JT ;
DEUTSCH, S ;
GESELOWITZ, DB ;
TARBELL, JM .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1994, 116 (02) :190-200
[5]   FACTORS INFLUENCING ERYTHROCYTE DESTRUCTION IN ARTIFICIAL ORGANS [J].
BERNSTEIN, EF ;
BLACKSHEAR, PL ;
KELLER, KH .
AMERICAN JOURNAL OF SURGERY, 1967, 114 (01) :126-+
[6]  
BLACKSHEAR PL, 1950, BIOMECHANICS ITS FDN, P501
[7]  
BLACKSHEAR PL, 1986, HDB BIOENGINEERING
[8]   3-DIMENSIONAL NUMERICAL PREDICTION OF STRESS LOADING OF BLOOD PARTICLES IN A CENTRIFUGAL PUMP [J].
BLUDSZUWEIT, C .
ARTIFICIAL ORGANS, 1995, 19 (07) :590-596
[9]  
BOROVETZ HS, 1989, CRIT REV BIOMED ENG, V17, P179
[10]   Computational fluid dynamics as a development tool for rotary blood pumps [J].
Burgreen, GW ;
Antaki, JF ;
Wu, ZJ ;
Holmes, AJ .
ARTIFICIAL ORGANS, 2001, 25 (05) :336-340