Comparison of continuous thermodilution and bolus cardiac output measurements in septic shock

被引:21
作者
Sun, QH
Rogiers, P
Pauwels, D
Vincent, JL
机构
[1] Free Univ Brussels, Erasme Univ Hosp, Dept Intens Care, B-1070 Brussels, Belgium
[2] Middelheim Hosp, Dept Intens Care, Antwerp, Belgium
关键词
hemodynamic measurement; monitoring system; fever; critically ill;
D O I
10.1007/s00134-002-1415-2
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: To compare continuous (CCO) and bolus (BCO) thermodilution cardiac output measurement techniques over a wide range of cardiac outputs and blood temperatures in a septic sheep model. Design and setting: Prospective experimental study in a university intensive care laboratory. Subjects: Thirty-five anesthetized sheep. Interventions: Pulmonary artery catheters allowing measurement of CCO and BCO were placed through the external jugular vein. Cecal ligation and perforation was performed to induce septic shock. In 14 sheep two femoral venous catheters were placed and connected to a hemofiltration system to alter blood temperature. Measurements: CCO and BCO were measured every hour during the experiment. Three 10-ml bolus injections of iced normal saline were given through a closed injectate system and then averaged . The CCO readings were collected just before the BCO measurements. The relationship between CCO and BCO was assessed using Bland and Altman's method. Results: In 465 paired data the temperature ranged between 34.0degrees and 40.9degreesC, CCO between 1.4 and 17.0 l/min, and BCO between 1.1 and 17.4 l/min. There was a highly significant correlation between CCO and BCO (r = 0.97). The bias (difference between CCO and BCO) was -0.19 l/min, the SD of the difference 0.45 l/min, and the limits of agreement -1.08/0.71 l/min. There were also highly significant C correlations between CCO and BCO at the different temperatures (extreme values: 34.0-34.9degreesC, r = 0.90; 40.0-40.9degreesC, r = 0.98). Conclusions: Thermodilution measurements of CCO are reliable, when compared to BCO measurements, over a large range of cardiac outputs and blood temperatures.
引用
收藏
页码:1276 / 1280
页数:5
相关论文
共 31 条
[1]   COMPARISON OF CARDIAC-OUTPUT MEASURED WITH CONTINUOUS THERMODILUTION AND CARDIAC-OUTPUT CALCULATED WITH THE FICKS PRINCIPLE [J].
BIZOUARN, P ;
BLANLOEIL, Y ;
PINAUD, M .
ANNALES FRANCAISES D ANESTHESIE ET DE REANIMATION, 1994, 13 (05) :685-689
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]  
BOLDT J, 1994, CRIT CARE MED, V22, P1913
[4]  
Bottiger BW, 1996, INTENS CARE MED, V22, P312
[5]   CONTINUOUS VERSUS INTERMITTENT CARDIAC-OUTPUT MEASUREMENT IN CARDIAC SURGICAL PATIENTS UNDERGOING HYPOTHERMIC CARDIOPULMONARY BYPASS [J].
BOTTIGER, BW ;
RAUCH, H ;
BOHRER, H ;
MOTSCH, J ;
SODER, M ;
FLEISCHER, F ;
MARTIN, E .
JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 1995, 9 (04) :405-411
[6]  
DAPER A, 1986, Acute Care, V12, P113
[7]  
DOBB GJ, 1987, INTENS CARE MED, V13, P304
[8]   CARDIAC-OUTPUT BY THERMODILUTION TECHNIQUE - EFFECT OF INJECTATES VOLUME AND TEMPERATURE ON ACCURACY AND REPRODUCIBILITY IN THE CRITICALLY ILL PATIENT [J].
ELKAYAM, U ;
BERKLEY, R ;
AZEN, S ;
WEBER, L ;
GEVA, B ;
HENRY, WL .
CHEST, 1983, 84 (04) :418-422
[9]   EVALUATION OF A NEW CONTINUOUS THERMODILUTION CARDIAC-OUTPUT MONITOR IN CRITICALLY ILL PATIENTS - A PROSPECTIVE CRITERION STANDARD STUDY [J].
HALLER, M ;
ZOLLNER, C ;
BRIEGEL, J ;
FORST, H .
CRITICAL CARE MEDICINE, 1995, 23 (05) :860-866
[10]  
Hogue C W Jr, 1994, J Cardiothorac Vasc Anesth, V8, P631, DOI 10.1016/1053-0770(94)90193-7