Clinical use of the Odstock dropped foot stimulator: Its effect on the speed and effort of walking

被引:182
作者
Taylor, PN [1 ]
Burridge, JH
Dunkerley, AL
Wood, DE
Norton, JA
Singleton, C
Swain, ID
机构
[1] Salisbury Dist Hosp, Dept Biomed Engn & Med Phys, Salisbury SP2 8BJ, Wilts, England
[2] City Hosp, Neurophysiotherapy Dept, Birmingham, W Midlands, England
[3] UCL, Dept Med Phys & Bioengn, Implanted Devices Grp, London, England
来源
ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION | 1999年 / 80卷 / 12期
关键词
D O I
10.1016/S0003-9993(99)90333-7
中图分类号
R49 [康复医学];
学科分类号
100215 ;
摘要
Objective: To assess the clinical effectiveness of the Odstock dropped foot stimulator by analysis of its effect on physiological cost index (PCI) and speed of walking, This functional electrical stimulation (FES) device stimulates the common peroneal nerve during the swing phase of gait. Design: A retrospective study of patients who had used the device for 4% months. Subjects: One hundred fifty-one patients with a dropped foot resulting from an upper motor neuron lesion. Setting: A medical physics and biomedical engineering department of a district general hospital specializing in the clinical application of FES and a neurophysiotherapy department at a separate hospital. Main Outcome Measures: Changes in walking speed and effort of walking, as measured by PCI over a 10-meter course. Results: There was a 92.7% compliance with treatment. Stroke patients showed a mean increase in walking speed of 27% (p < .01) and reduction in PCI of 31% (p < .01) with stimulation, and changes of 14% (p < .01) and 19% (p < .01), respectively, while not using the stimulator. Multiple sclerosis patients gained similar orthotic benefit but no "carry-over." Conclusions: The measured differences in walking with and without stimulation were statistically significant in the stroke and multiple sclerosis groups. In this study use of the stimulator improved walking, Those with stroke demonstrated a shortterm "carry-over" effect. (C) 1999 by the American Congress of Rehabilitation Medicine and the American Academy of Physical Medicine and Rehabilitation.
引用
收藏
页码:1577 / 1583
页数:7
相关论文
共 15 条
  • [1] [Anonymous], 1997, PHYSIOTHERAPY, DOI DOI 10.1016/S0031-9406(05)65582-4
  • [2] Bailey MJ., 1995, Physiotherapy, V81, P618
  • [3] Experience of clinical use of the Odstock dropped foot stimulator
    Burridge, J
    Taylor, P
    Hagan, S
    Swain, I
    [J]. ARTIFICIAL ORGANS, 1997, 21 (03) : 254 - 260
  • [4] The effects of common peroneal stimulation on the effort and speed of walking: a randomized controlled trial with chronic hemiplegic patients
    Burridge, JH
    Taylor, PN
    Hagan, SA
    Wood, DE
    Swain, ID
    [J]. CLINICAL REHABILITATION, 1997, 11 (03) : 201 - 210
  • [5] Peroneal stimulator: Evaluation for the correction of spastic drop foot in hemiplegia
    Granat, MH
    Maxwell, DJ
    Ferguson, ACB
    Lees, KR
    Barbenel, JC
    [J]. ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION, 1996, 77 (01): : 19 - 24
  • [6] Persisting symptoms and carers' views of outcome after subarachnoid haemorrhage
    Hellawell, DJ
    Taylor, R
    Pentland, B
    [J]. CLINICAL REHABILITATION, 1999, 13 (04) : 333 - 340
  • [7] KANDELL ER, 1991, CELLULAR MECH LEARNI, P1009
  • [8] OBJECTIVE QUANTIFICATION OF SPASTIC HYPERTONIA - CORRELATION WITH CLINICAL FINDINGS
    KATZ, RT
    ROVAI, GP
    BRAIT, C
    RYMER, WZ
    [J]. ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION, 1992, 73 (04): : 339 - 347
  • [9] Liberson W., 1961, ARCH PHYS MED REHAB, V42, P202
  • [10] Nene A., 1993, CLIN REHABIL, V7, P319, DOI DOI 10.1177/026921559300700408