The sevoflurane saving capacity of a new anaesthetic agent conserving device compared with a low flow circle system

被引:28
作者
Enlund, M
Lambert, H
Wiklund, L
机构
[1] Univ Uppsala, Dept Anaesthesia & Intens Care, Uppsala, Sweden
[2] Hudson RCI AB, Upplands Vasby, Sweden
关键词
equipment : breathing system; heat and moisture exchangers; vaporizers; anesthetics : volatile;
D O I
10.1034/j.1399-6576.2002.460506.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background: An anaesthetic agent conserving device (ACD) has been added to a Bain system to approach the agent-saving capacity of a low flow circle system. Methods: Randomly selected ASA physical status I patients received a standardized anaesthetic with sevoflurane in air/O-2 through either a circle system with 1.51/min fresh gas flow (n 8), or through a Bain system with an added ACD with fresh gas flow 4.4-6.41/min (n = 8). A target controlled infusion of remifentanil was used for analgesia. Results: The median sevoflurane consumption was 19.7 and 22.0ml/MAC/h with the low flow circle system and the Bain system + ACD, respectively (P=0.10, Mann-Whitney U-test), or when adjusted for weight 0.25 and 0.28ml/MAC/h/kg (P=0.26, Mann-Whitney U-test). Conclusion: The expenditure of sevoflurane with a Bain system + ACD was close to that in a circle system with 1.51/min fresh gas flow. It is thereby possible to use sevoflurane to all its potential, performing for example rapid alterations in end-tidal concentration using high fresh gas flows by combining a Bain system with an ACD. Although the price is not decided for this not yet commercially available device, a potential for a lower cost exists. Additionally, there will be no concerns of toxic compounds produced in the absorber.
引用
收藏
页码:506 / 511
页数:6
相关论文
共 16 条
[1]  
DALE O, 1991, ACTA ANAESTHESIO S96, V35, P40
[2]  
Eger EI, 2000, ANESTHESIA, P74
[3]   A new device to reduce the consumption of a halogenated anaesthetic agent [J].
Enlund, M ;
Wiklund, L ;
Lambert, H .
ANAESTHESIA, 2001, 56 (05) :429-432
[4]   CARBON-MONOXIDE PRODUCTION FROM DEGRADATION OF DESFLURANE, ENFLURANE, ISOFLURANE, HALOTHANE, AND SEVOFLURANE BY SODA LIME AND BARALYME(R) [J].
FANG, ZX ;
EGER, EI ;
LASTER, MJ ;
CHORTKOFF, BS ;
KANDEL, L ;
IONESCU, P .
ANESTHESIA AND ANALGESIA, 1995, 80 (06) :1187-1193
[5]   IS THE END-TIDAL PARTIAL-PRESSURE OF ISOFLURANE A GOOD PREDICTOR OF ITS ARTERIAL PARTIAL-PRESSURE [J].
FREI, FJ ;
ZBINDEN, AM ;
THOMSON, DA ;
RIEDER, HU .
BRITISH JOURNAL OF ANAESTHESIA, 1991, 66 (03) :331-339
[6]   Compound A uptake and metabolism to mercapturic acids and 3,3,3-trifluoro-2-fluoromethoxypropanoic acid during low-flow sevoflurane anesthesia - Biomarkers for exposure, risk assessment, and interspecies comparison [J].
Kharasch, ED ;
Jubert, C .
ANESTHESIOLOGY, 1999, 91 (05) :1267-1278
[7]   Reduction of isoflurane minimal alveolar concentration by remifentanil [J].
Lang, E ;
Kapila, A ;
Shlugman, D ;
Hoke, JF ;
Sebel, PS ;
Glass, PSA .
ANESTHESIOLOGY, 1996, 85 (04) :721-728
[8]  
Mapleson WW, 1996, BRIT J ANAESTH, V76, P179, DOI 10.1093/bja/76.2.179
[9]   Future applications for TCI systems [J].
Milne, SE ;
Kenny, GNC .
ANAESTHESIA, 1998, 53 :56-60
[10]  
OLIVEI M, 11 EUR C AN CENSA JU