Evaluation of a new generation of culture bottle using an automated bacterial culture system for detecting nine common contaminating organisms found in platelet components

被引:48
作者
Brecher, ME
Heath, DG
Hay, SN
Rothenberg, SJ
Stutzman, LC
机构
[1] Univ N Carolina, Dept Pathol & Lab Med, Transfus Med Serv, Chapel Hill, NC USA
[2] Univ N Carolina, Dept Clin Microbiol, Chapel Hill, NC USA
[3] Organon Teknika, BioMerieux, Durham, NC USA
关键词
D O I
10.1046/j.1537-2995.2002.00122.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND: An automated bacterial culture system (BacT/ALERT 3D, bioMerieux) has been previously validated with a variety of bacteria in platelets. The recovery of bacteria in platelets using a new generation of culture bottles that do not require venting and that use a liquid emulsion sensor was studied. STUDY DESIGN AND METHODS: Bacillus cereus, Enterobacter cloacae, Escherichia coli, Klebsiella oxytoca, Staphylococcus aureus, Staphylococcus epidermidis, Serratia marcescens, Streptococcus viridans, and Propionibacterium acnes isolates were inoculated into Day 2 platelets to concentrations of 10 and 100 CFU per mL. Samples were then studied with current and new aerobic, anaerobic, and pediatric bottles. RESULTS: All organisms, except P. acnes, were detected in a mean time of 9.2 to 20.4 (10 CFU/mL) or 8.7 to 18.6 (100 CFU/mL) hours. P. acnes was detected in a mean time of 69.2 (10 CFU/mL) or 66.0 (100 CFU/ mL) hours. The 10-fold increase in inoculum was associated with a mean 9.2 percent difference in detection time. The aerobic, anaerobic, and pediatric bottles had a mean difference in detection time (hours) between the current and new bottles of 0.10 (p = 0.61), 0.4 (p 0.38), and 1.0 (p < 0.001), respectively. CONCLUSION: No difference in detection time between the current and new aerobic and anaerobic bottles was demonstrated. The new pediatric bottles had a small but significant delay in detection.
引用
收藏
页码:774 / 779
页数:6
相关论文
共 23 条
  • [1] ESCHERICHIA-COLI SEPSIS FROM CONTAMINATED PLATELET TRANSFUSION
    ARNOW, PM
    WEISS, LM
    WEIL, D
    ROSEN, NR
    [J]. ARCHIVES OF INTERNAL MEDICINE, 1986, 146 (02) : 321 - 324
  • [2] AuBuchon JP, 2001, TRANSFUSION, V41, p1S
  • [3] BLAJCHMAN MA, 1997, TRANSFUSION S, V37, pS4
  • [4] Growth of bacteria in inoculated platelets: implications for bacteria detection and the extension of platelet storage
    Brecher, ME
    Holland, PV
    Pineda, AA
    Tegtmeier, GE
    Yomtovian, R
    [J]. TRANSFUSION, 2000, 40 (11) : 1308 - 1312
  • [5] Evaluation of an automated culture system for detecting bacterial contamination of platelets: an analysis with 15 contaminating organisms
    Brecher, ME
    Means, N
    Jere, CS
    Heath, D
    Rothenberg, S
    Stutzman, LC
    [J]. TRANSFUSION, 2001, 41 (04) : 477 - 482
  • [6] THE USE OF A CHEMILUMINESCENCE-LINKED UNIVERSAL BACTERIAL RIBOSOMAL-RNA GENE PROBE AND BLOOD-GAS ANALYSIS FOR THE RAPID DETECTION OF BACTERIAL-CONTAMINATION IN WHITE CELL-REDUCED AND NONREDUCED PLATELETS
    BRECHER, ME
    BOOTHE, G
    KERR, A
    [J]. TRANSFUSION, 1993, 33 (06) : 450 - 457
  • [7] DETECTION AND QUANTITATION OF BACTERIA IN PLATELET PRODUCTS STORED AT AMBIENT-TEMPERATURE
    BUCHHOLZ, DH
    YOUNG, VM
    FRIEDMAN, NR
    REILLY, JA
    MARDINEY, MR
    [J]. TRANSFUSION, 1973, 13 (05) : 268 - 275
  • [8] Cooper L, 1999, TRANSFUSION, V39, p119S
  • [9] BACTERIAL CONTAMINATION OF PLATELET CONCENTRATES STORED AT 20 DEGREES C
    CUNNINGHAM, M
    CASH, JD
    [J]. JOURNAL OF CLINICAL PATHOLOGY, 1973, 26 (06) : 401 - 404
  • [10] DODD RY, 1994, BLOOD SUPPLY RISKS P, P1