Chemometric quantitation of the active substance (containing CN) in a pharmaceutical tablet using near-infrared (NIR) transmittance and NIR FT-Raman spectra

被引:164
作者
Dyrby, M
Engelsen, SB
Norgaard, L
Bruhn, M
Lundsberg-Nielsen, L
机构
[1] Royal Vet & Agr Univ, Chemometr Grp, Dept Dairy & Food Sci, DK-1958 Frederiksberg C, Denmark
[2] H Lundbeck & Co AS, DK-2500 Valby, Denmark
关键词
chemometrics; content uniformity; cyanide; iPLS; near-infrared transmittance; NIR; pharmaceutical; PLS; quantitative; Raman; tablets;
D O I
10.1366/0003702021955358
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In this study, near-infrared (NIR) transmittance and Raman spectroscopy chemometric calibrations of the active substance content of a pharmaceutical tablet were developed using partial least-squares regression (PLS). Although the active substance contained the strongly Raman active Cequivalent toN functional group, the best results were obtained with NIR transmittance, which highlights the difference between (microscopic) surface sampling and whole tablet diffuse transmittance sampling. The tablets exist in four dosages with only two different concentrations of active substance (5 mg (5.6% w/w), and 10, 15, and 20 mg (8.0% w/w) active substance per tablet). A calibration on all four dosages resulted in a prediction error expressed as the root mean squared error of cross-validation (RMSECV) of 0.30% w/w for the NIR transmittance calibration. The corresponding error when using Raman spectra was 0-56% w/w. Specially prepared calibration batches covering the range 85-115% of the nominal content for each dosage were added to the first sample set, and NIR transmittance calibrations on this set-containing coated as well as uncoated tablets-gave a further reduction in prediction errors to 0.21-0.289% w/w. This corresponds to relative prediction errors (RMSECV/y(nom)) of 2.6-3.7%. This is a reasonable low error when compared to the error of the chromatographic reference method, which was estimated to 3.5%.
引用
收藏
页码:579 / 585
页数:7
相关论文
共 16 条
[1]  
[Anonymous], 1993, J NEAR INFRARED SPEC, DOI DOI 10.1255/jnirs.25
[2]   STANDARD NORMAL VARIATE TRANSFORMATION AND DE-TRENDING OF NEAR-INFRARED DIFFUSE REFLECTANCE SPECTRA [J].
BARNES, RJ ;
DHANOA, MS ;
LISTER, SJ .
APPLIED SPECTROSCOPY, 1989, 43 (05) :772-777
[3]  
CAMO ASA, 1998, UNSCR US MAN
[4]   Near infrared transmittance analysis for the assay of solid pharmaceutical dosage forms [J].
Corti, P ;
Ceramelli, G ;
Dreassi, E ;
Mattii, S .
ANALYST, 1999, 124 (05) :755-758
[5]   Rotating samples in FT-RAMAN spectrometers [J].
De Paepe, ATG ;
Dyke, JM ;
Hendra, PJ ;
Langkilde, FW .
SPECTROCHIMICA ACTA PART A-MOLECULAR AND BIOMOLECULAR SPECTROSCOPY, 1997, 53 (13) :2261-2266
[6]   Determination of paracetamol in intact tablets by use of near infrared transmittance spectroscopy [J].
Eustaquio, A ;
Blanco, M ;
Jee, RD ;
Moffat, AC .
ANALYTICA CHIMICA ACTA, 1999, 383 (03) :283-290
[7]   Vibrational spectrometry for the assessment of active substance in metoprolol tablets: A comparison between transmission and diffuse reflectance near-infrared spectrometry [J].
Gottfries, J ;
Depui, H ;
Fransson, M ;
Jongeneelen, M ;
Josefson, M ;
Langkilde, FW ;
Witte, DT .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1996, 14 (11) :1495-1503
[8]  
Grummisch U, 1998, PHARM IND, V60, P1002
[9]   QUANTITATIVE-ANALYSIS USING RAMAN METHODS [J].
JAWHARI, T ;
HENDRA, PJ ;
WILLIS, HA ;
JUDKINS, M .
SPECTROCHIMICA ACTA PART A-MOLECULAR AND BIOMOLECULAR SPECTROSCOPY, 1990, 46 (02) :161-170
[10]  
Jedvert I, 1998, JOURNAL OF NEAR INFRARED SPECTROSCOPY, VOL 6 1998, P279