Quantification of the anti-leukemia drug ST1571 (Gleevec™) and its metabolite (CGP 74588) in monkey plasma using a semi-automated solid phase extraction procedure and liquid chromatography-tandem mass spectrometry

被引:54
作者
Bakhtiar, R [1 ]
Khemani, L [1 ]
Hayes, M [1 ]
Bedman, T [1 ]
Tse, F [1 ]
机构
[1] Novartis Inst Biomed Res, Dept Drug Metab & Pharmacokinet, E Hanover, NJ 07936 USA
关键词
Gleevec (TM); STI571; leukemia; Bcr-Abl; CGP; 74588; imatinib mesylate;
D O I
10.1016/S0731-7085(02)00080-8
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Signal Transduction Inhibitor 571 (STI571, formerly known as CGP 57148B) or Gleevec(TM) received fast track approval by the US Food and Drug Administration (FDA) for treatment of chronic myeloid leukemia (CML). STI571 (Gleevec(TM)) is a revolutionary and promising new oral therapy for CML, which functions at the molecular level with high specificity. The dramatic improvement in efficacy compared with existing treatments prompted an equally profound increase in the pace of development of Gleevec(TM). The duration from first dose in man to completion of the Now Drug Application (NDA) filing was less than 3 years. In addition, recently, FDA approved Gleevec(TM) for the treatment of gastrointestinal stromal tumor (GIST). In order to support all toxicokinetic (TK) studies with sufficient speed to meet various target dates, a semi-automated procedure using solid phase extraction (SPE) was developed and validated. A Packard Multi-Probe I and a SPE step in a 96-well plate format were utilized. A 3M Empore octyl (Q-standard density 96-well plate was used for plasma sample extraction. A Sciex API 3000 triple quadrupole mass spectrometer with an atmospheric pressure chemical ionization (APCI) interface operated in positive ion mode was used for detection. Lower limits of quantification of 1.00 and 2.00 ng/ml were attained for STI571 and its metabolite, CGP 74588, respectively. The method proved to be rugged and allowed the simultaneous quantification of STI571 and CGP 74588 in monkey plasma. Herein, assay development, validation, and representative concentration-time profiles obtained from TK studies are presented. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:1183 / 1194
页数:12
相关论文
共 49 条