Automatic analysis of auditory nerve electrically evoked compound action potential with an artificial neural network

被引:6
作者
Charasse, B
Thai-Van, H
Chanal, JM
Berger-Vachon, C
Collet, L
机构
[1] Lab Neurosci & Syst Sensoriels, CNRS, UMR 5020, F-69366 Lyon, France
[2] Hop Edouard Herriot, Serv Audiol & Eplorat Orofaciales, F-69003 Lyon, France
关键词
artificial neural network; neural response telemetry (NRT); pattern recognition; threshold estimation; cochlear implant; amplitude growth function; auditory nerve; electrically evoked compound action potential;
D O I
10.1016/j.artmed.2004.03.004
中图分类号
TP18 [人工智能理论];
学科分类号
081104 ; 0812 ; 0835 ; 1405 ;
摘要
The auditory nerve's electrically evoked compound action potential is recorded in deaf patients equipped with the Nucleus' 24 cochlear implant using a reverse telemetry system (NRT(TM)). Since the threshold of the NRT response (NRT-T) is thought to reflect the psychophysics needed for programming cochlear implants, efforts have been made by specialized management teams to develop its use. This study aimed at developing a valid tool based on artificial neural networks (ANN) technology, for automatic estimation of NRT-T. The ANN used was a single layer perceptron, trained with 120 NRT traces. Learning traces differed from data used for the validation. A total of 550 NRT traces from 11 cochlear implant subjects were analyzed separately by the! system and by a group of physicians with expertise in NRT analysis. Both worked to determine 37 NRT-T values, using the response amplitude growth function (AGF) (linear regression of response amplitudes obtained at decreasing stimulus intensity levels). The validity of the system was assessed by comparing the NRT-T values automatically determined by the system with those determined by the physicians. A strong correlation was found between automatic and physician-obtained NRT-T values (Pearson r correlation coefficient >0.9). ANOVA statistics confirmed that automatic NRT-Ts did not differ from physician-obtained values (F = 0.08999, P = 0.03). Moreover, the average error between NRT-Ts predicted by the system and NRT-Ts measured by the physicians (3.6 stimulation units) did not differ significantly from the average error between NRT-Ts measured by each of the three physicians (4.2 stimulation units). In conclusion, the automatic system developed in this study was found to be as efficient as human experts for fitting the amplitude growth function and estimating NRT-T, with the advantage of considerable time-saving. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:221 / 229
页数:9
相关论文
共 24 条
[1]   Summary of results using the nucleus CI24M implant to record the electrically evoked compound action potential [J].
Abbas, PJ ;
Brown, CJ ;
Shallop, JK ;
Firszt, JB ;
Hughes, ML ;
Hong, SH ;
Staller, SJ .
EAR AND HEARING, 1999, 20 (01) :45-59
[2]  
[Anonymous], P IEEE INT C NEUR NE
[3]  
BISHOP CM, 1995, NEURAL NETWORKS PATT, P95
[4]   ELECTRICALLY EVOKED WHOLE-NERVE ACTION-POTENTIALS - PARAMETRIC DATA FROM THE CAT [J].
BROWN, CJ ;
ABBAS, PJ .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1990, 88 (05) :2205-2210
[5]   ELECTRICALLY EVOKED WHOLE-NERVE ACTION-POTENTIALS - DATA FROM HUMAN COCHLEAR IMPLANT USERS [J].
BROWN, CJ ;
ABBAS, PJ ;
GANTZ, B .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1990, 88 (03) :1385-1391
[6]  
Brown CJ, 1998, AM J OTOL, V19, P320
[7]   The relationship between EAP and EABR thresholds and levels used to program the nucleus 24 speech processor: Data from adults [J].
Brown, CJ ;
Hughes, ML ;
Luk, B ;
Abbas, PJ ;
Wolaver, A ;
Gervais, J .
EAR AND HEARING, 2000, 21 (02) :151-163
[8]   A prognostic computer model to individually predict post-procedural complications in interventional cardiology -: The INTERVENT Project [J].
Budde, T ;
Haude, M ;
Höpp, HW ;
Kerber, S ;
Caspari, G ;
Fassbender, G ;
Fingerhut, M ;
Novopashenny, I ;
Ogurol, Y ;
Breithardt, G ;
Erbel, R ;
Erdmann, E ;
Wischnewsky, MB .
EUROPEAN HEART JOURNAL, 1999, 20 (05) :354-363
[9]  
CHARASSE C, 2004, IN PRESS INT J AUDIO
[10]   Preliminary neural response telemetry results [J].
Cullington, H .
BRITISH JOURNAL OF AUDIOLOGY, 2000, 34 (03) :131-140