Potentiometric detection of carboxylic acids by flow injection analysis using a tungsten oxide electrode

被引:14
作者
Chen, ZL
Alexander, PW
机构
[1] Department of Physical Science, Univ. of Tasmania at Launceston, Launceston, Tasmania 7250
关键词
flow injection; carboxylic acids; tungsten oxide electrode; potentiometry;
D O I
10.1016/0003-2670(96)00258-9
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Potentiometric flow injection (PFI) analysis of carboxylic acids based on a tungsten oxide electrode is described. The electrode is shown to offer high sensitivity and a stable baseline for the detection of carboxylic acids. The parameters affecting the electrode response such as the concentration of buffer, the pH and flow rate were studied in detail. The electrode exhibited a linear response for propionic acid in the concentration range of 2.5x10(-4)M-5x10(-5)M using a 5x10(-4)M phosphate buffer at pH 6.5 as a carrier. The detection limits were found to be in the range of 1.0x10(-5)M-5x10(-5)M with a 20 mu l injection.
引用
收藏
页码:187 / 192
页数:6
相关论文
共 16 条
[1]   CONTINUOUS-FLOW POTENTIOMETRIC MONITORING OF ALPHA-AMINO-ACIDS WITH COPPER WIRE AND TUBULAR ELECTRODES [J].
ALEXANDER, PW ;
MAITRA, C .
ANALYTICAL CHEMISTRY, 1981, 53 (11) :1590-1594
[2]   POTENTIOMETRIC FLOW-INJECTION DETERMINATION OF COPPER COMPLEXING INORGANIC ANIONS WITH A COPPER WIRE INDICATOR ELECTRODE [J].
ALEXANDER, PW ;
HADDAD, PR ;
TROJANOWICZ, M .
ANALYTICAL CHEMISTRY, 1984, 56 (13) :2417-2422
[3]   INDIRECT POTENTIOMETRIC DETERMINATION OF METAL-IONS BY FLOW-INJECTION ANALYSIS WITH A COPPER ELECTRODE [J].
ALEXANDER, PW ;
TROJANOWICZ, M ;
HADDAD, PR .
ANALYTICAL LETTERS PART A-CHEMICAL ANALYSIS, 1984, 17 (04) :309-320
[4]   ONLINE DETERMINATION OF ACETIC-ACID IN A CONTINUOUS PRODUCTION OF ACETOBACTER-ACETICUS [J].
BECKER, T ;
KITTSTEINEREBERLE, R ;
LUCK, T ;
SCHMIDT, HL .
JOURNAL OF BIOTECHNOLOGY, 1993, 31 (03) :267-275
[5]   FLOW-INJECTION ANALYSIS WITH ELECTROCHEMICAL DETECTION [J].
CAMMANN, K .
FRESENIUS ZEITSCHRIFT FUR ANALYTISCHE CHEMIE, 1988, 329 (06) :691-697
[6]   ENHANCEMENT OF SENSOR SELECTIVITY BY GAS-DIFFUSION SEPARATION .1. FLOW-INJECTION POTENTIOMETRIC DETERMINATION OF CYANIDE WITH A METALLIC SILVER-WIRE ELECTRODE [J].
FRENZEL, W ;
LIU, CY ;
OLEKSYFRENZEL, J .
ANALYTICA CHIMICA ACTA, 1990, 233 (01) :77-84
[7]  
FRITZ JS, 1987, QUANTITATIVE ANAL CH, P123
[8]  
GLOD BK, 1995, J CHROMATOGR A, V699, P31
[9]   EFFECT OF ADDITION OF MAIN ION TO CARRIER SOLUTION IN POTENTIOMETRIC FLOW-INJECTION MEASUREMENTS WITH SOLID-STATE ION-SELECTIVE ELECTRODES [J].
ILCHEVA, L ;
TROJANOWICZ, M ;
KRAWCZYK, TKV .
FRESENIUS ZEITSCHRIFT FUR ANALYTISCHE CHEMIE, 1987, 328 (1-2) :27-32
[10]   Ion-selective electrodes in organic analysis: Determination of carboxylic acids via in situ conversion into amines [J].
Lee, AWM ;
Chan, WH ;
Lam, YS .
ANALYST, 1995, 120 (12) :2841-2844