摘要
The globalisation of the analysis of a series of individual measurements often results in more robust and reliable outcomes. However, instrumental drifts that can occur between individual measurements destroy the ideal data structure and thus the advantages. A method based on rank annihilation factor analysis (RAFA) is introduced for the correction of several types of instrumental inconsistencies. It can be applied to many series of bilinear datasets. Experimental examples discussed in this paper comprise the successful correction of non-uniform retention time drifts in chromatography due to temperature or pressure changes, wavelength shifts in IR spectroscopy in an industrial control situation, and background absorption shifts in UV-VIS spectroscopy applied to equilibrium investigations. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:249 / 259
页数:11
相关论文
共 23 条
[1]   SIMPLEX FOCUSING OF RETENTION TIMES AND LATENT VARIABLE PROJECTIONS OF CHROMATOGRAPHIC PROFILES [J].
ANDERSSON, R ;
HAMALAINEN, MD .
CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 1994, 22 (01) :49-61
[2]  
[Anonymous], 1990, Journal ofChemometrics, DOI DOI 10.1002/CEM.1180040105
[3]   PARAFAC. Tutorial and applications [J].
Bro, R .
CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 1997, 38 (02) :149-171
[4]   2ND-ORDER GLOBALIZATION FOR THE DETERMINATION OF ACTIVATION PARAMETERS IN KINETICS [J].
BUGNON, P ;
CHOTTARD, JC ;
JESTIN, JL ;
JUNG, B ;
LAURENCZY, G ;
MAEDER, M ;
MERBACH, AE ;
ZUBERBUHLER, AD .
ANALYTICA CHIMICA ACTA, 1994, 298 (02) :193-201
[5]   MATHEMATICAL-ANALYSIS OF NORMALIZATION TECHNIQUES USED IN CHROMATOGRAPHY [J].
CHILCOTE, DD ;
SCOTT, CD .
ANALYTICAL CHEMISTRY, 1973, 45 (04) :721-724
[6]   Second order global analysis: the evaluation of series of spectrophotometric titrations for improved determination of equilibrium constants [J].
Dyson, RM ;
Kaderli, S ;
Lawrance, GA ;
Maeder, M ;
Zunderbuhler, AD .
ANALYTICA CHIMICA ACTA, 1997, 353 (2-3) :381-393
[7]   COMPUTERIZED DATA ACQUISITION AND DATA REDUCTION IN SPECTROPHOTOMETRIC ANALYSIS OF METAL-ION COMPLEXES .1. GENERAL-CONSIDERATIONS AND DATA ACQUISITION [J].
GAMPP, H ;
MAEDER, M ;
ZUBERBUHLER, AD .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 1988, 7 (03) :111-113
[8]   Correction of mass spectral drift using artificial neural networks [J].
Goodacre, R ;
Kell, DB .
ANALYTICAL CHEMISTRY, 1996, 68 (02) :271-280
[9]   RETENTION TIME SHIFT ADJUSTMENTS OF 2-WAY CHROMATOGRAMS USING BESSELS INEQUALITY [J].
GRUNG, B ;
KVALHEIM, OM .
ANALYTICA CHIMICA ACTA, 1995, 304 (01) :57-66
[10]   APPLICATION OF METHOD OF RANK ANNIHILATION TO QUANTITATIVE-ANALYSES OF MULTICOMPONENT FLUORESCENCE DATA FROM VIDEO FLUOROMETER [J].
HO, CN ;
CHRISTIAN, GD ;
DAVIDSON, ER .
ANALYTICAL CHEMISTRY, 1978, 50 (08) :1108-1113