Rank annihilation correction for the amendment of instrumental inconsistencies

被引:23
作者
Maeder, M
Neuhold, YM
Olsen, A
Puxty, G
Dyson, R
Zilian, A
机构
[1] Univ Newcastle, Dept Chem, Callaghan, NSW 2308, Australia
[2] Solvias AG, Online Technol, CH-4002 Basel, Switzerland
关键词
correction; bilinear data; rank annihilation; chromatography; baseline;
D O I
10.1016/S0003-2670(02)00488-9
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The globalisation of the analysis of a series of individual measurements often results in more robust and reliable outcomes. However, instrumental drifts that can occur between individual measurements destroy the ideal data structure and thus the advantages. A method based on rank annihilation factor analysis (RAFA) is introduced for the correction of several types of instrumental inconsistencies. It can be applied to many series of bilinear datasets. Experimental examples discussed in this paper comprise the successful correction of non-uniform retention time drifts in chromatography due to temperature or pressure changes, wavelength shifts in IR spectroscopy in an industrial control situation, and background absorption shifts in UV-VIS spectroscopy applied to equilibrium investigations. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:249 / 259
页数:11
相关论文
共 23 条
[21]   MULTICOMPONENT DETERMINATION OF CHLORINATED HYDROCARBONS USING A REACTION-BASED CHEMICAL SENSOR .3. MEDIUM-RANK 2ND-ORDER CALIBRATION WITH RESTRICTED TUCKER MODELS [J].
SMILDE, AK ;
TAULER, R ;
HENSHAW, JM ;
BURGESS, LW ;
KOWALSKI, BR .
ANALYTICAL CHEMISTRY, 1994, 66 (20) :3345-3351
[22]   STANDARDIZATION OF 2ND-ORDER INSTRUMENTS [J].
WANG, YD ;
KOWALSKI, BR .
ANALYTICAL CHEMISTRY, 1993, 65 (09) :1174-1180
[23]   Automatic correction of peak shifts in Raman spectra before PLS regression [J].
Witjes, H ;
van den Brink, M ;
Melssen, WJ ;
Buydens, LMC .
CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 2000, 52 (01) :105-116