Practical assessment of frictional heating effects and thermostat design on the performance of conventional (3 μm and 5 μm) columns in reversed-phase high-performance liquid chromatography

被引:51
作者
Fallas, Morgane M. [1 ]
Hadley, Mark R. [2 ]
McCalley, David V. [1 ]
机构
[1] Univ W England, Biomed Res Ctr, Bristol BS16 1QY, Avon, England
[2] AstraZeneca, Global Proc R&D, Macclesfield SK10 2NA, Cheshire, England
关键词
HPLC; Frictional heating; Retention; Column efficiency; HIGH-PRESSURE; HIGH-TEMPERATURE; SELECTIVITY DIFFERENCES; ELEVATED-TEMPERATURES; ID COLUMNS; EFFICIENCY; PARTICLES; SEPARATIONS; PREDICTION; RESOLUTION;
D O I
10.1016/j.chroma.2009.03.007
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A practical investigation of frictional heating effects in conventional C18 columns was undertaken, to investigate whether problems found for sub-2 mu m columns were also present for those of particle size 3 mu m and 5 mu m and different internal diameter. The influence of a water bath, a still air heater, and a forced air heater on performance was investigated. Heating effects were substantial, with a decrease in k of almost 15% for toluene over the flow rate range similar to 0.4-2.3 mL/min with a 15 cm x 0.46cm ID column packed with 3 mu m particles. Heating effects on retention increased with increasing solute k,with increase in the column ID, with decrease in the column particle size, and with decrease in the set column oven temperature. While the water bath minimised axial temperature gradients and thus its effect on k, radial temperature gradients were potentially serious with this system, especially at high mobile phase velocity, even with columns containing 5 mu m particles. In contrast to the effects of axial temperature gradients in 4.6 mm columns, very little difference in Van Deemter plots was noted between the three different thermostats with 2 mm ID columns, even when 3 mu m particles were used. However, the efficiency of 2 mm columns for peaks of low or moderate k (k<4) can be compromised by the extra dead volume introduced by the heating systems, even with conventional HPLC systems with otherwise minimised extra column volume. (C) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:3961 / 3969
页数:9
相关论文
共 37 条
[1]  
ANTIA FD, 1988, J CHROMATOGR, V435, P1
[2]   Rationalisation of unusual changes in efficiency and retention with temperature shown for bases in reversed-phase high-performance liquid chromatography at intermediate pH [J].
Buckenmaier, SMC ;
McCalley, DV ;
Euerby, MR .
JOURNAL OF CHROMATOGRAPHY A, 2004, 1060 (1-2) :117-126
[3]  
Chen H., 1993, Analytical Methods and Instrumentation, V1, P213
[4]  
Chen MH, 1997, J CHROMATOGR A, V788, P51
[5]   Very high pressure HPLC with 1 mm id columns [J].
Colón, LA ;
Cintrón, JM ;
Anspach, JA ;
Fermier, AM ;
Swinney, KA .
ANALYST, 2004, 129 (06) :503-504
[6]   Fused-core particle technology as an alternative to sub-2-αm particles to achieve high separation efficiency with low backpressure [J].
Cunliffe, Jennifer M. ;
Maloney, Todd D. .
JOURNAL OF SEPARATION SCIENCE, 2007, 30 (18) :3104-3109
[7]   Influence of frictional heating on temperature gradients in ultra-high-pressure liquid chromatography on 2.1 mm ID columns [J].
de Villiers, A ;
Lauer, H ;
Szucs, R ;
Goodall, S ;
Sandra, P .
JOURNAL OF CHROMATOGRAPHY A, 2006, 1113 (1-2) :84-91
[8]   Selectivity differences for C18 and C8 reversed-phase columns as a function of temperature and gradient steepness I.: Optimizing selectivity and resolution [J].
Dolan, JW ;
Snyder, LR ;
Blanc, T ;
Van Heukelem, L .
JOURNAL OF CHROMATOGRAPHY A, 2000, 897 (1-2) :37-50
[9]   Selectivity differences for C18 and C8 reversed-phase columns as a function of temperature and gradient steepness II.: Minimizing column reproducibility problems [J].
Dolan, JW ;
Snyder, LR ;
Blanc, T .
JOURNAL OF CHROMATOGRAPHY A, 2000, 897 (1-2) :51-63
[10]   Temperature selectivity in reversed-phase high performance liquid chromatography [J].
Dolan, JW .
JOURNAL OF CHROMATOGRAPHY A, 2002, 965 (1-2) :195-205