MAHOROBA, first-in-man study: 6-month results of a biodegradable polymer sustained release tacrolimus-eluting stent in de novo coronary stenoses

被引:20
作者
Onuma, Yoshinobu [1 ]
Serruys, Patrick [1 ]
den Heijer, Peter [2 ]
Joesoef, Kaiyum Sheik [3 ]
Duckers, Henricus [1 ]
Regar, Evelyn [1 ]
Kukreja, Neville [1 ]
Tanimoto, Shuzou [1 ]
Garcia-Garcia, Hector M. [4 ]
van Beusekom, Heleen [1 ]
van der Giessen, Willem [1 ]
Nishide, Takuji [5 ]
机构
[1] Erasmus MC, Thoraxctr, NL-3015 CE Rotterdam, Netherlands
[2] AMPHIA Hosp, Breda, Netherlands
[3] Med Ctr Rijnmond Zuid, Rotterdam, Netherlands
[4] Cardialysis BV, Rotterdam, Netherlands
[5] Kaneka Corp, Osaka, Japan
关键词
Tacrolimus-eluting stent; First-in-man study; Drug-eluting stent; Coronary artery disease; TISSUE FACTOR EXPRESSION; WIKTOR STENT; QUANTITATIVE-ANALYSIS; ULTRASOUND IMAGES; SMOOTH-MUSCLE; TAXUS-II; SIROLIMUS; IMPLANTATION; PROLIFERATION; RESTENOSIS;
D O I
10.1093/eurheartj/ehp127
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
To report the 4-month angiographic and 6-month clinical follow-up in first-in-man study using the tacrolimus-eluting bioabsorbable polymer-coated cobalt-chromium MAHOROBA (TM) stent. A total of 47 patients with either stable angina or unstable angina, or silent myocardial ischaemia, based on a de novo coronary stenosis that could be covered by a single 18 mm stent in a native coronary artery with a diameter between 3.0 and 3.5 mm were enrolled at three sites. The primary endpoint was in-stent late loss at 4 months. The secondary endpoints include %volume obstruction of the stents assessed by intravascular ultrasound (IVUS) at 4 months and major adverse cardiac events (MACE) at 6 months. Forty-seven patients were enrolled. Procedural success was achieved in 97.9%. At 4-month follow-up, in-stent late loss was 0.99 +/- 0.46 mm, whereas in-stent %volume obstruction in IVUS was 34.8 +/- 15.8%. At 6 months, there were no deaths, but 2 patients suffered from a myocardial infarction and 11 patients required ischaemia-driven repeat revascularization. The composite MACE rate was 23.4%. This tacrolimus-eluting stent failed to prevent neointimal hyperplasia, despite the theoretical advantages of the tacrolimus, which has less inhibitory effects on endothelial cells than smooth muscle cells.
引用
收藏
页码:1477 / 1485
页数:9
相关论文
共 36 条
[1]   Peristent remodeling and neointimal suppression 2 years after polymer-based, paclitaxel-eluting stent implantation - Insights from serial intravascular ultrasound analysis in the TAXUS II study [J].
Aoki, J ;
Colombo, A ;
Dudek, D ;
Banning, AP ;
Drzewiecki, J ;
Zmudka, K ;
Schiele, F ;
Russell, ME ;
Koglin, J ;
Serruys, PW .
CIRCULATION, 2005, 112 (25) :3876-3883
[2]  
BARTORELLI A, 2004, SORIN JANUS FIM RESU
[3]   Clinical end points in coronary stent trials - A case for standardized definitions [J].
Cutlip, Donald E. ;
Windecker, Stephan ;
Mehran, Roxana ;
Boam, Ashley ;
Cohen, David J. ;
van Es, Gerrit-Anne ;
Steg, P. Gabriel ;
Morel, Marie-angele ;
Mauri, Laura ;
Vranckx, Pascal ;
McFadden, Eugene ;
Lansky, Alexandra ;
Hamon, Martial ;
Krucoff, Mitchell W. ;
Serruys, Patrick W. .
CIRCULATION, 2007, 115 (17) :2344-2351
[4]  
DAWKINS KD, 2008, RANDOMIZED MULTICENT
[5]   Retrospective image-based gating of intracoronary ultrasound images for improved quantitative analysis: The intelligate method [J].
de Winter, SA ;
Hamers, R ;
Degertekin, M ;
Tanabe, K ;
Lemos, PA ;
Serruys, PW ;
Roelandt, JRTC ;
Bruining, N .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2004, 61 (01) :84-94
[6]   Intravascular ultrasound evaluation after sirolimus eluting stent implantation for de novo and in-stent restenosis lesions [J].
Degertekin, M ;
Lemos, PA ;
Lee, CH ;
Tanabe, K ;
Sousa, JE ;
Abizaid, A ;
Regar, E ;
Sianos, G ;
van der Giessen, WJ ;
de Feyter, PJ ;
Wuelfert, E ;
Popma, JJ ;
Serruys, PW .
EUROPEAN HEART JOURNAL, 2004, 25 (01) :32-38
[7]   ANGIOGRAPHIC PREDICTORS OF RECURRENCE OF RESTENOSIS AFTER WIKTOR STENT IMPLANTATION IN NATIVE CORONARY-ARTERIES [J].
DEJAEGERE, P ;
SERRUYS, PW ;
BERTRAND, M ;
WIEGAND, V ;
MARQUIS, JF ;
VROLICX, M ;
PIESSENS, J ;
VALEIX, B ;
KOBER, G ;
BONNIER, H ;
RUTSCH, W ;
UEBIS, R .
AMERICAN JOURNAL OF CARDIOLOGY, 1993, 72 (02) :165-170
[8]  
DEJAEGERE P, 1994, CATHETER CARDIO DIAG, V32, P147
[9]   Pathological mechanisms of fatal late coronary stent thrombosis in humans [J].
Farb, A ;
Burke, AP ;
Kolodgie, FD ;
Virmani, R .
CIRCULATION, 2003, 108 (14) :1701-1706
[10]   Pathological correlates of late drug-eluting stent thrombosis - Strut coverage as a marker of endothelialization [J].
Finn, Aloke V. ;
Joner, Michael ;
Nakazawa, Gaku ;
Kolodgie, Frank ;
Newell, John ;
John, Mike C. ;
Gold, Herman K. ;
Virmani, Renu .
CIRCULATION, 2007, 115 (18) :2435-2441