Long-term hardware-related complications of deep brain stimulation

被引:18
作者
Oh, MY [1 ]
Abosch, A [1 ]
Kim, SH [1 ]
Lang, AE [1 ]
Lozano, AM [1 ]
机构
[1] Univ Toronto, Toronto Hosp, Western Div, Dept Surg,Div Neurosurg, Toronto, ON M5T 2S8, Canada
关键词
adverse events; deep brain stimulation; hardware;
D O I
暂无
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
OBJECTIVE: To determine the incidence of long-term hardware-related complications of deep brain stimulation (DBS). METHODS: The study design is a retrospective chart review of a single-surgeon, single-institution experience with DBS in 84 consecutive cases from 1993 to 1999. Only patients with a minimum follow-up of I year were considered. Five patients were excluded because trial stimulation failed to achieve pain relief (n = 4) or because the procedure was aborted owing to hemorrhage (n = 1). Seventy-nine patients received 124 permanent DBS electrode implants. RESULTS: The mean follow-up period was 33 months, and the cumulative follow-up time was 217 patient-years or 310 electrode-years. Overall, 20 patients (25.3%) had 26 hardware-related complications involving 23 (18.5%) of the electrodes. There were 4 lead fractures, 4 lead migrations, 3 short or open circuits, 12 erosions and/or infections, 2 foreign body reactions, and one cerebrospinal fluid leak. The hardware-related complication rate per electrode-year was 8.4%. The most common complications were related to the electrode connectors. A significant finding was a high number of complications involving erosions or infections, which occurred in 7 of 12 instances as a late complication (beyond 12 mo). CONCLUSION: Long-term follow-up reveals that hardware-related complications occur in a significant number of patients. Factors that lead to such complications must be identified and addressed to maximize the important benefits of DBS therapy.
引用
收藏
页码:1268 / 1274
页数:7
相关论文
共 13 条
[1]
MOTOR EFFECTS OF STIMULATING THE HUMAN CEREBELLAR THALAMUS [J].
ASHBY, P ;
LANG, AE ;
LOZANO, AM ;
DOSTROVSKY, JO .
JOURNAL OF PHYSIOLOGY-LONDON, 1995, 489 (01) :287-298
[2]
Benabid AL, 1998, MOVEMENT DISORD, V13, P119
[3]
BENDOK B, 1998, TXB STEREOTACTIC FUN, P01539
[4]
Anchoring of deep brain stimulation electrodes using a microplate [J].
Favre, J ;
Taha, JM ;
Steel, T ;
Burchiel, KJ .
JOURNAL OF NEUROSURGERY, 1996, 85 (06) :1181-1183
[5]
High-frequency unilateral thalamic stimulation in the treatment of essential and parkinsonian tremor [J].
Koller, W ;
Pahwa, R ;
Busenbark, K ;
Hubble, J ;
Wilkinson, S ;
Lang, A ;
Tuite, P ;
Sime, E ;
Lazano, A ;
Hauser, R ;
Malapira, T ;
Smith, D ;
Tarsy, D ;
Miyawaki, E ;
Norregaard, T ;
Kormos, T ;
Olanow, CW .
ANNALS OF NEUROLOGY, 1997, 42 (03) :292-299
[6]
Deep brain stimulation for intractable pain: A 15-year experience [J].
Kumar, K ;
Toth, C ;
Nath, RK .
NEUROSURGERY, 1997, 40 (04) :736-746
[7]
TREATMENT OF CHRONIC PAIN BY DEEP BRAIN-STIMULATION - LONG-TERM FOLLOW-UP AND REVIEW OF THE LITERATURE [J].
LEVY, RM ;
LAMB, S ;
ADAMS, JE .
NEUROSURGERY, 1987, 21 (06) :885-893
[8]
Multicentre European study of thalamic stimulation in parkinsonian and essential tremor [J].
Limousin, P ;
Speelman, JD ;
Gielen, F ;
Janssens, M .
JOURNAL OF NEUROLOGY NEUROSURGERY AND PSYCHIATRY, 1999, 66 (03) :289-296
[9]
*MEDTR INC, 2000, DEEP BRAIN STIM 3387
[10]
Oh MY, 2001, NEUROSURGERY, V49, P363