Rapid confirmation/quantitation of cocaine and benzoylecgonine in urine utilizing high performance liquid chromatography and tandem mass spectrometry

被引:47
作者
Jeanville, PM
Estapé, ES
Needham, SR
Cole, MJ
机构
[1] Pfizer Inc, Div Cent Res, Drug Metab Technol Dept, Groton, CT 06340 USA
[2] Univ Puerto Rico, Coll Hlth Related Profess, San Juan, PR 00936 USA
关键词
D O I
10.1016/S1044-0305(99)00138-5
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A rapid, but sensitive and selective method for confirmation and quantitation of benzoylecgonine (BZE) and cocaine (COC) in urine by fast-gradient liquid chromatography/tandem mass spectrometry (LC/MS/MS) is described. The chromatographic separation was performed on a reversed phase column employing fast-gradient techniques. Matrix prepared standards, blanks, and QC's were filtered then aliquots were transferred into a 96 well plate. Injection volumes of 25 mu L were made onto the analytical column, with the flow diverted from the atmospheric pressure ionization source for the first 0.5 min of the analysis. Simultaneous multiple reaction monitoring (MRM) of three discrete transitions for each compound were used to identify BZE and COG. Quantitation was achieved utilizing the most prominent parent-daughter transition and internal standard calibration techniques. The coefficients of variation (CV) for the analysis of these drugs ranged from 0.6% to 6.8% at a concentration of 150 ng/mL (n = 155). This method suggests that fast-gradient LC/MS/MS may be suitable for routine confirmation of immunoassay cocaine-positive samples. (C) 2000 American Society for Mass Spectrometry.
引用
收藏
页码:257 / 263
页数:7
相关论文
共 21 条
[1]   SERIAL PLASMA-CONCENTRATIONS OF COCAETHYLENE, COCAINE, AND ETHANOL IN TRAUMA VICTIMS [J].
BAILEY, DN .
JOURNAL OF ANALYTICAL TOXICOLOGY, 1993, 17 (02) :79-83
[2]   Determination of common drugs of abuse in body fluids using one isolation procedure and liquid chromatography atmospheric pressure chemical ionization mass spectrometry [J].
Bogusz, MJ ;
Maier, RD ;
Kruger, KD ;
Kohls, U .
JOURNAL OF ANALYTICAL TOXICOLOGY, 1998, 22 (07) :549-558
[3]  
BRAVO B, 1997, 14 P M INT ASS FOR S, P208
[4]   Narrow-bore HPLC in combination with fluorescence and electrospray mass spectrometric detection for the analysis of cocaine and metabolites in human hair [J].
Clauwaert, KM ;
Van Bocxlaer, JF ;
Lambert, WE ;
Van den Eeckhout, EG ;
Lemière, F ;
Esmans, EL ;
De Leenheer, AP .
ANALYTICAL CHEMISTRY, 1998, 70 (11) :2336-2344
[5]   A multiple-site laboratory evaluation of three on-site urinalysis drug-testing devices [J].
Crouch, DJ ;
Frank, JF ;
Farrell, LJ ;
Karsch, HM ;
Klaunig, JE .
JOURNAL OF ANALYTICAL TOXICOLOGY, 1998, 22 (06) :493-502
[6]  
*DEPT HHS, 1994, FED REGISTER, P29908
[7]  
ENKE CG, UNPUB
[8]  
JOHANSSON M, 1996, J CHROMT A, V729, P279
[9]   UTILIZATION OF DIRECT LIQUID INLET LC-MS IN STUDIES OF PHARMACOLOGICAL AND TOXICOLOGICAL IMPORTANCE [J].
KENYON, CN ;
MELERA, A ;
ERNI, F .
JOURNAL OF ANALYTICAL TOXICOLOGY, 1981, 5 (05) :216-230
[10]  
MCCURDY HH, 1998, J ANAL TOXICOL, V12, P260