A linear, separable two-parameter model for dual energy CT imaging of proton stopping power computation

被引:66
作者
Han, Dong [1 ]
Siebers, Jeffrey V. [2 ]
Williamson, Jeffrey F. [1 ]
机构
[1] Virginia Commonwealth Univ, Dept Radiat Oncol, Med Phys Grad Program, Med Coll Virginia Campus, Richmond, VA 23298 USA
[2] Univ Virginia, Dept Radiat Oncol, Charlottesville, VA 22908 USA
关键词
dual energy; computed tomography; proton therapy; stopping power; HOUNSFIELD UNITS; HUMAN TISSUES; BODY-TISSUES; BONE-MARROW; DENSITY; CALIBRATION; SINGLE; NUMBERS;
D O I
10.1118/1.4939082
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
Purpose: To evaluate the accuracy and robustness of a simple, linear, separable, two-parameter model (basis vector model, BVM) in mapping proton stopping powers via dual energy computed tomography (DECT) imaging. Methods: The BVM assumes that photon cross sections (attenuation coefficients) of unknown materials are linear combinations of the corresponding radiological quantities of dissimilar basis substances (i.e., polystyrene, CaCl2 aqueous solution, and water). The authors have extended this approach to the estimation of electron density and mean excitation energy, which are required parameters for computing proton stopping powers via the Bethe-Bloch equation. The authors compared the stopping power estimation accuracy of the BVM with that of a nonlinear, nonseparable photon cross section Torikoshi parametric fit model (VCU tPFM) as implemented by the authors and by Yang et al. ["Theoretical variance analysis of single-and dual-energy computed tomography methods for calculating proton stopping power ratios of biological tissues," Phys. Med. Biol. 55, 1343-1362 (2010)]. Using an idealized monoenergetic DECT imaging model, proton ranges estimated by the BVM, VCU tPFM, and Yang tPFM were compared to International Commission on Radiation Units and Measurements (ICRU) published reference values. The robustness of the stopping power prediction accuracy of tissue composition variations was assessed for both of the BVM and VCU tPFM. The sensitivity of accuracy to CT image uncertainty was also evaluated. Results: Based on the authors' idealized, error-free DECT imaging model, the root-mean-square error of BVM proton stopping power estimation for 175 MeV protons relative to ICRU reference values for 34 ICRU standard tissues is 0.20%, compared to 0.23% and 0.68% for the Yang and VCU tPFM models, respectively. The range estimation errors were less than 1 mm for the BVM and Yang tPFM models, respectively. The BVM estimation accuracy is not dependent on tissue type and proton energy range. The BVM is slightly more vulnerable to CT image intensity uncertainties than the tPFM models. Both the BVM and tPFM prediction accuracies were robust to uncertainties of tissue composition and independent of the choice of reference values. This reported accuracy does not include the impacts of I-value uncertainties and imaging artifacts and may not be achievable on current clinical CT scanners. Conclusions: The proton stopping power estimation accuracy of the proposed linear, separable BVM model is comparable to or better than that of the nonseparable tPFM models proposed by other groups. In contrast to the tPFM, the BVM does not require an iterative solving for effective atomic number and electron density at every voxel; this improves the computational efficiency of DECT imaging when iterative, model-based image reconstruction algorithms are used to minimize noise and systematic imaging artifacts of CT images. (C) 2016 American Association of Physicists in Medicine.
引用
收藏
页码:600 / 612
页数:13
相关论文
共 42 条
[21]
ICRU, 1992, report 46
[22]
ICRU, 1984, Technical Report 37
[23]
Material differentiation by dual energy CT: initial experience [J].
Johnson, Thorsten R. C. ;
Krauss, Bernhard ;
Sedlmair, Martin ;
Grasruck, Michael ;
Bruder, Herbert ;
Morhard, Dominik ;
Fink, Christian ;
Weckbach, Sabine ;
Lenhard, Miriam ;
Schmidt, Bernhard ;
Flohr, Thomas ;
Reiser, Maximilian F. ;
Becker, Christoph R. .
EUROPEAN RADIOLOGY, 2007, 17 (06) :1510-1517
[24]
Deriving effective atomic numbers from DECT based on a parameterization of the ratio of high and low linear attenuation coefficients [J].
Landry, Guillaume ;
Seco, Joao ;
Gaudreault, Mathieu ;
Verhaegen, Frank .
PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (19) :6851-6866
[25]
Deriving concentrations of oxygen and carbon in human tissues using single- and dual-energy CT for ion therapy applications [J].
Landry, Guillaume ;
Parodi, Katia ;
Wildberger, Joachim E. ;
Verhaegen, Frank .
PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (15) :5029-5048
[26]
CALIBRATION AND RESPONSE OF AN EMI SCANNER [J].
RUTHERFORD, RA ;
PULLAN, BR ;
ISHERWOOD, I .
NEURORADIOLOGY, 1976, 11 (01) :7-13
[27]
MEASUREMENT OF EFFECTIVE ATOMIC NUMBER AND ELECTRON-DENSITY USING AN EMI SCANNER [J].
RUTHERFORD, RA ;
PULLAN, BR ;
ISHERWOOD, I .
NEURORADIOLOGY, 1976, 11 (01) :15-21
[29]
Experimental feasibility of multi-energy photon-counting K-edge imaging in pre-clinical computed tomography [J].
Schlomka, J. P. ;
Roessl, E. ;
Dorscheid, R. ;
Dill, S. ;
Martens, G. ;
Istel, T. ;
Baeumer, C. ;
Herrmann, C. ;
Steadman, R. ;
Zeitler, G. ;
Livne, A. ;
Proksa, R. .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (15) :4031-4047
[30]
Patient specific optimization of the relation between CT-Hounsfield units and proton stopping power with proton radiography [J].
Schneider, U ;
Pemler, P ;
Besserer, J ;
Pedroni, E ;
Lomax, A ;
Kaser-Hotz, B .
MEDICAL PHYSICS, 2005, 32 (01) :195-199