Safety of orthopedic implants in magnetic resonance imaging: An experimental verification

被引:58
作者
Kumar, Ritabh [1 ]
Lerski, Richard A. [1 ]
Gandy, Stephen [1 ]
Clift, Benedict A. [1 ]
Abboud, Rami J. [1 ]
机构
[1] Univ Dundee, Ninewells Hosp & Med Sch, TORT Ctr,Div Surg & Oncol, IMAR, Dundee DD1 9SY, Scotland
关键词
MRI; orthopedic implants; safety; ferromagnetism; heating;
D O I
10.1002/jor.20213
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Magnetic resonance imaging (MRI) is an indispensable tool for musculoskeletal imaging., The presence of metal, however, raises concerns. The potential risks are loosening and migration of the implant, heating of the metal with surrounding tissue, causing thermal damage, and artifactual distortion which compromise the diagnostic value of the procedure. The aim of this study was to test experimentally the nature and extent of the first two of these effects in orthopedic implants. The degree of ferromagnetism was evaluated by deflection studies at the portals of a 0.25 Tesla permanent magnet and 1.0 Tesla clinical MRI scanner. None of the orthopedic implants exhibited any attraction. External fixator clamps, however, showed significant ferromagnetism. The heating of implants by "worst-case" scenario imaging sequences was insignificant. Many contemporary nonferromagnetic orthopedic implants can be imaged safely. It is prudent, however, to perform ex vivo deflection studies on a duplicate implant to confirm MR compatibility. With external fixator devices exhibiting strong ferromagnetism, MRI should be avoided. With expanding indications for MRI, orthopedic implants are unlikely to limit the potential of this powerful. tool. (c) 2006 Orthopaedic Research Society.
引用
收藏
页码:1799 / 1802
页数:4
相关论文
共 13 条
[1]
POTENTIAL HAZARDS IN NMR IMAGING - HEATING EFFECTS OF CHANGING MAGNETIC-FIELDS AND RF FIELDS ON SMALL METALLIC IMPLANTS [J].
DAVIS, PL ;
CROOKS, L ;
ARAKAWA, M ;
MCREE, R ;
KAUFMAN, L ;
MARGULIS, AR .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1981, 137 (04) :857-860
[2]
DISEGI JA, 1992, INJURY, V23, P1
[3]
MR IMAGING IN PATIENTS WITH METALLIC IMPLANTS [J].
LAAKMAN, RW ;
KAUFMAN, B ;
HAN, JS ;
NELSON, AD ;
CLAMPITT, M ;
OBLOCK, AM ;
HAAGA, JR ;
ALFIDI, RJ .
RADIOLOGY, 1985, 157 (03) :711-714
[4]
POTENTIAL HAZARDS AND ARTIFACTS OF FERROMAGNETIC AND NONFERROMAGNETIC SURGICAL AND DENTAL MATERIALS AND DEVICES IN NUCLEAR MAGNETIC-RESONANCE IMAGING [J].
NEW, PFJ ;
ROSEN, BR ;
BRADY, TJ ;
BUONANNO, FS ;
KISTLER, JP ;
BURT, CT ;
HINSHAW, WS ;
NEWHOUSE, JH ;
POHOST, GM ;
TAVERAS, JM .
RADIOLOGY, 1983, 147 (01) :139-148
[5]
Perren SM, 1991, MANUAL INTERNAL FIXA
[6]
RUPP R, 1993, SPINE, V18, P379
[7]
MR IMAGING AND METALLIC IMPLANTS FOR ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION - ASSESSMENT OF FERROMAGNETISM AND ARTIFACT [J].
SHELLOCK, FG ;
MINK, JH ;
CURTIN, S ;
FRIEDMAN, MJ .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1992, 2 (02) :225-228
[8]
HIGH-FIELD-STRENGTH MR IMAGING AND METALLIC BIOMEDICAL IMPLANTS - AN EXVIVO EVALUATION OF DEFLECTION FORCES [J].
SHELLOCK, FG ;
CRUES, JV .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1988, 151 (02) :389-392
[9]
MR PROCEDURES AND BIOMEDICAL IMPLANTS, MATERIALS, AND DEVICES - 1993 UPDATE [J].
SHELLOCK, FG ;
MORISOLI, S ;
KANAL, E .
RADIOLOGY, 1993, 189 (02) :587-599
[10]
Biomedical implants and devices: Assessment of magnetic field interactions with a 3.0-Tesla MR system [J].
Shellock, FG .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2002, 16 (06) :721-732