Popping phenomena in temperature-controlled radiofrequency ablation: When and why do they occur?

被引:57
作者
Eick, OJ
Gerritse, B
Schumacher, B
机构
[1] Medtron Bakken Res Ctr, NL-6229 GW Maastricht, Netherlands
[2] Univ Bonn, D-5300 Bonn, Germany
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2000年 / 23卷 / 02期
关键词
catheter ablation; radiofrequency; flow; electrode contact; popping;
D O I
10.1111/j.1540-8159.2000.tb00807.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
During temperature-controlled radiofrequency (RF) ablation a popping sound sometimes occurs. This popping phenomenon is know to be associated with unwanted effects like blood boiling, endocardial rupture, catheter dislocation, and impedance rise. The present in vitro study determined the influence of cooling, electrode contact, and tip temperature on the occurrence of popping phenomena. Pieces of porcine ventricle were immersed in a bath of saline solution at 37 degrees C. Forty-two RF ablations were performed with different electrode-tissue contact forces (i.e., 0.0-0.44N) in a temperature-controlled mode (70 degrees C setpoint, 30 s, 50 W maximum polyer output, 4-mm tip, thermocouple). Half of the 42 ablations were performed with fluid flow (0.1 m/s, group I), the other half without flow (group nl. In group I, mean tip temperature and power were 55.6 +/- 8.5 degrees C and 36.2 +/- 13.8 W,resulting in a lesion volume of 121 +/- 57 mm(3). In group II, the respective values were 67.3 +/- 1.5 degrees C and 9.9 +/- 5.2 W resulting in a volume of 42 +/- 18 mm(3). The differences between groups were statistically significant. Overall, ten popping phenomenas occurred in group I and none in group II. Pops occurred significantly more often when the contact force was < 0.1 N (8/10) and the tip temperature was < 60 degrees C (8/10). Two endocardial ruptures occurred, both were associated with a popping phenomenon. Using temperature control, the probability of pops is significantly high er when the ablation electrode and the endocardial tissue surface are exposed to fluid flow and the electrode-tissue contact is poor. Under these conditions the tissue temperature can be much higher than the temperature measured at the tip electrode and can potentially reach 100 degrees C causing intramyocardial steam formation and a popping phenomenon.
引用
收藏
页码:253 / 258
页数:6
相关论文
共 12 条
[1]   Radiofrequency current ablation of porcine right atrium: Increased lesion size with bipolar two catheter technique compared to unipolar application in vitro and in vivo [J].
Anfinsen, OG ;
Kongsgaard, E ;
Foerster, A ;
Aass, H ;
Amlie, JP .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (01) :69-78
[2]   The effects of electrode-tissue contact on radiofrequency lesion generation [J].
Avitall, B ;
Mughal, K ;
Hare, J ;
Helms, R ;
Krum, D .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1997, 20 (12) :2899-2910
[3]  
AVITALL B, 1992, CIRCULATION, P86
[4]   The LETR-Principle: A novel method to assess electrode-tissue contact in radiofrequency ablation [J].
Eick, OJ ;
Wittkampf, FH ;
Bronneberg, T ;
Schumacher, B .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1998, 9 (11) :1180-1185
[5]  
HAINES D E, 1991, Journal of Cardiovascular Electrophysiology, V2, P509, DOI 10.1111/j.1540-8167.1991.tb01353.x
[6]   OBSERVATIONS ON ELECTRODE-TISSUE INTERFACE TEMPERATURE AND EFFECT ON ELECTRICAL-IMPEDANCE DURING RADIOFREQUENCY ABLATION OF VENTRICULAR MYOCARDIUM [J].
HAINES, DE ;
VEROW, AF .
CIRCULATION, 1990, 82 (03) :1034-1038
[7]   SIGNIFICANCE OF PHYSICAL PARAMETERS FOR THE EFFICACY OF RADIOFREQUENCY CATHETER ABLATION [J].
HOFFMANN, E ;
DORWARTH, U ;
PULTER, R ;
GOKEL, M ;
STEINBECK, G .
BIOMEDIZINISCHE TECHNIK, 1992, 37 (04) :62-68
[8]   BIOPHYSICAL PARAMETERS OF RADIOFREQUENCY CATHETER ABLATION [J].
HOFFMANN, E ;
HABERL, R ;
PULTER, R ;
GOKEL, M ;
STEINBECK, G .
INTERNATIONAL JOURNAL OF CARDIOLOGY, 1992, 37 (02) :213-222
[9]   EFFECTS OF COMBINED RADIOFREQUENCY AND DIRECT-CURRENT ENERGY CATHETER ABLATION ON VENTRICULAR MYOCARDIUM IN PIGS [J].
KONGSGAARD, E ;
FOERSTER, A ;
AASS, H ;
MADSEN, S ;
AMLIE, JP .
EUROPEAN HEART JOURNAL, 1995, 16 (04) :514-520
[10]   Inverse relationship between electrode size and lesion size during radiofrequency ablation with active electrode cooling [J].
Nakagawa, H ;
Wittkampf, FHM ;
Yamanashi, WS ;
Pitha, JV ;
Imai, S ;
Campbell, B ;
Arruda, M ;
Lazzara, R ;
Jackman, WM .
CIRCULATION, 1998, 98 (05) :458-465