Evaluation of matrix-assisted laser desorption ionization-time-of-flight mass measurement accuracy by using delayed extraction

被引:92
作者
Edmondson, RD [1 ]
Russell, DH [1 ]
机构
[1] TEXAS A&M UNIV,DEPT CHEM,LAB BIOL MASS SPECTROMETRY,COLLEGE STN,TX 77843
基金
美国国家科学基金会;
关键词
D O I
10.1016/1044-0305(96)00027-X
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Mass measurement accuracy of a reflectron time-of-flight mass spectrometer equipped with delayed extraction is evaluated. Mass resolution of 10,000-15,000 is achieved routinely for peptides in the mass range of 1-6 ku ionized by matrix-assisted laser desorption ionization. The mass measurement accuracy of peptides with molecular weights of 1-4 ku is 10-15 ppm by using external calibration and better than 5 ppm by using internal calibration. The mass calibration remains accurate over a broad mass-to-charge ratio range (1-4 ku), even when the calibration curve is extrapolated several thousand mass units.
引用
收藏
页码:995 / 1001
页数:7
相关论文
共 25 条