Feasibility and safety of a novel technology for pacing without leads

被引:48
作者
Echt, Debra S. [1 ]
Cowan, Mark W. [1 ]
Riley, Richard E. [1 ]
Brisken, Axel F. [1 ]
机构
[1] EBR Syst Inc, Sunnyvale, CA 94085 USA
关键词
leadless pacing; electrical stimulation; new technotogy; ultrasound;
D O I
10.1016/j.hrthm.2006.06.012
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND Pacemaker lead complications and failures remain clinical problems. New devices incorporating three leads are associated with even greater limitations. OBJECTIVES The purpose of this study was to investigate the feasibility and safety of a technology enabling cardiac pacing without leads in an acute porcine model. METHODS The system is composed of an ultrasound transmitter delivering energy from the chest wall to a receiver-electrode in contact with the myocardium that then converts the ultrasound energy to electrical energy sufficient to pace. In five feasibility studies, the receiver-electrodes were attached to the tip of a catheter to facilitate intracardiac positioning at pacing sites. In six safety studies (five treatment and one sham), ultrasound energy was transmitted to both chest walls, and histopathologic examinations were performed to evaluate bioeffects due to ultrasound energy transmission. RESULTS In five feasibility studies, direct and ultrasound-mediated electrical pacing was demonstrated at 30 sites in the right atrium, right ventricle, and left ventricle, at direct electrical pacing outputs of 1.4 +/- 0.6 V and ultrasound-mediated electrical pacing outputs of 1.8 +/- 0.9 V. The mechanical index was 0.6 +/- 0.4 at the receiver site during ultrasound-mediated pacing at a depth of 11.2 +/- 2.4 cm from the chest wall. Using two receiver-electrode catheters, biventricular pacing was demonstrated in all studies. In five safety study treatment animals at a similar depth, the peak mechanical index was 2.3, and the thermal index was 0.4. Microscopic evaluation revealed no evidence of mechanical or thermal bioeffects. CONCLUSION The feasibility and safety of this novel technology for pacing without leads has been demonstrated acutely in animals.
引用
收藏
页码:1202 / 1206
页数:5
相关论文
共 21 条
[1]   Rationale and derivation of Mi and Ti - A review [J].
Abbott, JG .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1999, 25 (03) :431-441
[2]   Biophysical aspects of diagnostic ultrasound [J].
Barnett, SB .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2000, 26 :S68-S70
[3]   Complications of biventricular pacing [J].
Bhatta, L ;
Luck, JC ;
Wolbrette, DL ;
Naccarelli, GV .
CURRENT OPINION IN CARDIOLOGY, 2004, 19 (01) :31-35
[4]   Complications due to abandoned noninfected pacemaker leads [J].
Böhm, A ;
Pintér, A ;
Duray, G ;
Lehoczky, D ;
Dudás, G ;
Tomcsányi, I ;
Préda, I .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2001, 24 (12) :1721-1724
[5]   Triple-site pacing in patients with biventricular device - Incidence of the phenomenon and cardiac resynchronization benefit [J].
Bulava, A ;
Ansalone, G ;
Ricci, R ;
Giannantoni, P ;
Pignalberi, C ;
Heinc, P ;
Lukl, J ;
Santini, M .
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2004, 10 (01) :37-45
[6]   The search for cavitation in vivo [J].
Carstensen, EL ;
Gracewski, S ;
Dalecki, D .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2000, 26 (09) :1377-1385
[7]   Mechanical bioeffects of ultrasound [J].
Dalecki, D .
ANNUAL REVIEW OF BIOMEDICAL ENGINEERING, 2004, 6 :229-248
[8]   Acute Hemodynamic benefits of bi-atrial atrioventricular sequential pacing with the optimal atrioventricular delay [J].
Doi, A ;
Takagi, M ;
Toda, I ;
Yoshiyama, M ;
Takeuchi, K ;
Yoshikawa, J .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2005, 46 (02) :320-326
[9]   Delayed complications following pacemaker implantation [J].
Ellenbogen, KA ;
Wood, MA ;
Shepard, RK .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2002, 25 (08) :1155-1158
[10]  
FDA, 1997, INF MAN SEEK MARK CL