Assessment of a theoretical formalism for dose estimation in CT: an anthropomorphic phantom study

被引:87
作者
Brix, G [1 ]
Lechel, U
Veit, R
Truckenbrodt, R
Stamm, G
Coppenrath, EM
Griebel, J
Nagel, HD
机构
[1] Fed Office Radiat Protect, Dept Radiat Protect & Hlth, Div Med Radiat Hyg & Dosimetry, D-85764 Neuherberg, Germany
[2] Hannover Med Sch, Dept Radiol, D-3000 Hannover, Germany
[3] Univ Munich, Dept Radiol, Munich, Germany
[4] Philips Med Syst, Dept Sci & Technol, Hamburg, Germany
关键词
computed tomography; patient exposure; dosimetry; anthropomorphic phantom; scanner matching;
D O I
10.1007/s00330-004-2267-7
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Dose assessment in computed tomography (CT) is challenging due to the vast variety of CT scanners and imaging protocols in use. In the present study, the accurateness of a theoretical formalism implemented in the PC program CT-EXPO for dose calculation was evaluated by means of phantom measurements. Phantom measurements were performed with four I-slice, four 4-slice and two 16-slice spiral CT scanners. Firstly, scanner-specific nCTDIw values were measured and compared with the corresponding standard values used for dose calculation. Secondly, effective doses were determined for three CT scans (head, chest and pelvis) performed at each of the ten installations from readings of thermoluminescent dosimeters distributed inside an anthropomorphic Alderson phantom and compared with the corresponding dose values computed with CT-EXPO. Differences between standard and individually measured nCTDIw values were less than 16%. Statistical analysis yielded a highly significant correlation (P<0.001) between calculated and measured effective doses. The systematic and random uncertainty of the dose values calculated using standard nCTDIw values was about -9 and +/-11%, respectively. The phantom measurements and model calculations were carried out for a variety of CT scanners and representative scan protocols validate the reliability of the dosimetric formalism considered-at least for patients with a standard body size and a tube voltage of 120 kV selected for the majority of CT scans performed in our study.
引用
收藏
页码:1275 / 1284
页数:10
相关论文
共 35 条
  • [1] [Anonymous], 1982, S885 GSF
  • [2] [Anonymous], 1999, 16262 EUR EN
  • [3] Multidetector-array CT: Once again, technology creates new opportunities
    Berland, LL
    Smith, JK
    [J]. RADIOLOGY, 1998, 209 (02) : 327 - 329
  • [4] Radiation exposure in multi-slice versus single-slice spiral CT: results of a nationwide survey
    Brix, G
    Nagel, HD
    Stamm, G
    Veit, R
    Lechel, U
    Griebel, J
    Galanski, M
    [J]. EUROPEAN RADIOLOGY, 2003, 13 (08) : 1979 - 1991
  • [5] Effective doses in standard protocols for multi-slice CT scanning
    Cohnen, M
    Poll, LW
    Puettmann, C
    Ewen, K
    Saleh, A
    Mödder, U
    [J]. EUROPEAN RADIOLOGY, 2003, 13 (05) : 1148 - 1153
  • [6] Multi-slice technology in computed tomography
    Dawson, P
    Lees, WR
    [J]. CLINICAL RADIOLOGY, 2001, 56 (04) : 302 - 309
  • [7] Radiation dose in helical CT for detection of pulmonary embolism
    Diederich, S
    [J]. EUROPEAN RADIOLOGY, 2003, 13 (07) : 1491 - 1493
  • [8] Low-dose multislice CT of the thorax in follow-up of malignant lymphoma and extrapulmonary primary tumors
    Dinkel, HP
    Sonnenschein, M
    Hoppe, H
    Vock, P
    [J]. EUROPEAN RADIOLOGY, 2003, 13 (06) : 1241 - 1249
  • [9] European Commission, 2000, 19604 EUR EN
  • [10] Computed tomography practice Germany - Results of a national survey in 1999 - Preface
    Galanski, M
    Nagel, HD
    Stamm, G
    [J]. ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2001, 173 (10): : R1 - R1