A new method for attenuation coefficient measurement in the liver - Comparison with the spectral shift central frequency method

被引:72
作者
Fujii, Y [1 ]
Taniguchi, N
Itoh, K
Shigeta, K
Wang, Y
Tsao, JW
Kumasaki, K
Itoh, T
机构
[1] Jichi Med Sch, Dept Clin Lab Med, Minami Kawachi, Tochigi 3290498, Japan
[2] Aloka co Ltd, Tokyo, Japan
关键词
attenuation coefficient; cirrhotic liver; fatty liver; tissue characterization; ultrasonography;
D O I
10.7863/jum.2002.21.7.783
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Objective. To assess the feasibility of a new method of measuring the attenuation coefficient in the liver, which offers less variability of results than the conventional method. Methods. The attenuation coefficient was evaluated on the basis of the following equation with sound field correction: alpha = 1/4[f(c)(z)-epsilon(z)] . d/dz [log(M-0(z))-tau(z)]. In our system, the attenuation coefficient was also evaluated by the spectral shift central frequency method at the same time. We used 44 cases of normal liver, 40 cases of fatty liver, and 20 cases of cirrhotic liver in the system. Results. With this new method, attenuation coefficient values were 0.59 +/- 0.10 dB (.) cm(-1) (.) MHz(-1) in normal livers, 0.80 +/- 0.12 dB (.) cm(-1) (.) MHz(-1) in fatty livers, and 0.62 +/- 0.09 dB (.) cm(-1) (.) MHz(-1) in cirrhotic livers. In both methods we recorded a statistically significant difference between normal and fatty livers and between fatty and cirrhotic livers (P<.0001). Only in the fatty liver was any significant difference (P<.0001) found between attenuation coefficients in the new method and those in the spectral shift central frequency method (0.70 +/- 0.05 dB (.) cm(-1) (.) MHz(-1)). Conclusions. This new method, which was more sensitive in detecting fatty infiltration than the spectral shift central frequency method, was considered usable for evaluating the attenuation coefficient of the liver in vivo.
引用
收藏
页码:783 / 788
页数:6
相关论文
共 23 条
[1]   SPECKLE IN ULTRASOUND B-MODE SCANS [J].
BURCKHARDT, CB .
IEEE TRANSACTIONS ON SONICS AND ULTRASONICS, 1978, 25 (01) :1-6
[2]   ULTRASONIC SIGNAL-PROCESSING FOR INVIVO ATTENUATION MEASUREMENT - SHORT-TIME FOURIER-ANALYSIS [J].
FINK, M ;
HOTTIER, F ;
CARDOSO, JF .
ULTRASONIC IMAGING, 1983, 5 (02) :117-135
[3]   DIFFRACTION EFFECTS IN PULSE-ECHO MEASUREMENT [J].
FINK, MA ;
CARDOSO, JF .
IEEE TRANSACTIONS ON SONICS AND ULTRASONICS, 1984, 31 (04) :313-329
[4]   ULTRASOUND ATTENUATION MEASUREMENTS OF THE LIVER INVIVO USING A COMMERCIAL SECTOR SCANNER [J].
GARRA, BS ;
SHAWKER, TH ;
NASSI, M ;
RUSSELL, MA .
ULTRASONIC IMAGING, 1984, 6 (04) :396-407
[5]  
GOSS S A, 1979, Ultrasound in Medicine and Biology, V5, P181, DOI 10.1016/0301-5629(79)90086-3
[6]   ATTENUATION ESTIMATION ON PHANTOMS - A STABILITY TEST [J].
HE, P ;
GREENLEAF, JF .
ULTRASONIC IMAGING, 1986, 8 (01) :1-10
[7]   STUDIES ON FREQUENCY-DEPENDENT ATTENUATION IN THE NORMAL LIVER AND SPLEEN AND IN LIVER-DISEASES, USING THE SPECTRAL-SHIFT ZERO-CROSSING METHOD [J].
ITOH, K ;
YASUDA, Y ;
SUZUKI, O ;
ITOH, H ;
ITOH, T ;
TSAO, JW ;
KONISHI, T ;
KOYANO, A .
JOURNAL OF CLINICAL ULTRASOUND, 1988, 16 (08) :553-562
[8]   CLINICAL-APPLICATION OF AN ULTRASOUND ATTENUATION COEFFICIENT ESTIMATION TECHNIQUE FOR LIVER PATHOLOGY CHARACTERIZATION [J].
KUC, R .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1980, 27 (06) :312-319
[9]  
KUC R, 1978, IEEE ULTRASON S P, V26, P353
[10]   Ultrasound backscatter and attenuation in human liver with diffuse disease [J].
Lu, ZF ;
Zagzebski, JA ;
Lee, FT .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1999, 25 (07) :1047-1054