Differential interferences of hemoglobin and hemolysis on insulin assay with the Abbott Architect®-Ci8200 immunoassay

被引:15
作者
Garinet, Simon [1 ]
Fellahi, Soraya [1 ,2 ,3 ]
Marlin, Ginette [1 ]
Capeau, Jacqueline [1 ,2 ,3 ]
Lefevre, Guillaume [1 ,2 ,3 ]
Bastard, Jean-Philippe [1 ,2 ,3 ]
机构
[1] Hop Tenon, AP HP, Dept Biochem, F-75020 Paris, France
[2] Univ Paris 06, INSERM, Inst Cardiometab & Nutr, ICAN,U938, F-75012 Paris, France
[3] Univ Paris 06, UMR S 938, F-75005 Paris, France
关键词
Insulin; Hemolysis; Hemoglobin; SERUM;
D O I
10.1016/j.clinbiochem.2014.01.026
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives: This paper evaluates the effect of hemoglobin (Hb) and hemolysis on insulin measurements with the Architect (R)-Ci8200 analyzer. Design and methods: Insulin concentrations were measured using the Architect (R)-Ci8200. Interference studies were performed by spiking serum pools of defined insulin concentrations with increasing concentrations of either free Hb or hemolysate. A change of more than 10% was taken as evidence of significant interference. Results: A significant negative bias in insulin results was observed only in samples spiked with hemolysate but not with free Hb. The bias was proportional to the degree of hemolysis and to the time elapsed before insulin assay. This interference was decreased when samples were kept at + 4 degrees C. Conclusions: For all insulin requests, hemolysis must be systematically checked before biological interpretation of insulin results. (C) 2014 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:445 / 447
页数:3
相关论文
共 10 条
[1]   How can we measure insulin sensitivity/resistance? [J].
Antuna-Puente, B. ;
Disse, E. ;
Rabasa-Lhoret, R. ;
Laville, M. ;
Capeau, J. ;
Bastard, J. -P. .
DIABETES & METABOLISM, 2011, 37 (03) :179-188
[2]  
Chevenne D, 1998, CLIN CHEM, V44, P354
[3]   Effect of hemolysis on insulin determination by the Beckman Coulter Unicell DXI 800 immunoassay analyzer [J].
Cook, Paul R. ;
Glenn, Christine ;
Armston, Annie .
CLINICAL BIOCHEMISTRY, 2010, 43 (06) :621-622
[4]   Best Practice No 173 - Clinical and laboratory investigation of adult spontaneous hypoglycaemia [J].
Gama, R ;
Teale, JD ;
Marks, V .
JOURNAL OF CLINICAL PATHOLOGY, 2003, 56 (09) :641-646
[5]   Measurement of the high-molecular weight form of adiponectin in plasma is useful for the prediction of insulin resistance and metabolic syndrome [J].
Hara, Kazuo ;
Horikoshi, Momoko ;
Yamauchi, Toshimasa ;
Yago, Hirokazu ;
Miyazaki, Osamu ;
Ebinuma, Hiroyuki ;
Imai, Yasushi ;
Nagai, Ryozo ;
Kadowaki, Takashi .
DIABETES CARE, 2006, 29 (06) :1357-1362
[6]  
KROLL MH, 1994, CLIN CHEM, V40, P1996
[7]   Hemolyzed specimens: a major challenge for emergency departments and clinical laboratories [J].
Lippi, Giuseppe ;
Plebani, Mario ;
Di Somma, Salvatore ;
Cervellin, Gianfranco .
CRITICAL REVIEWS IN CLINICAL LABORATORY SCIENCES, 2011, 48 (03) :143-153
[8]   Performance evaluation and cross-reactivity from insulin analogs with the ARCHITECT insulin assay [J].
Moriyama, Masako ;
Hayashi, Nobuhide ;
Ohyabu, Chinami ;
Mukai, Masahiko ;
Kawano, Seiji ;
Kumagai, Shunichi .
CLINICAL CHEMISTRY, 2006, 52 (07) :1423-1426
[9]   Insulin measurements in haemolysed serum: influence of insulinase inhibitors [J].
Sapin, R ;
Ongagna, JC ;
Gasser, F ;
Grucker, D .
CLINICA CHIMICA ACTA, 1998, 274 (01) :111-117
[10]  
Vermeer HJ, 2005, CLIN CHEM, V51, P244