Detection limit of isotope dilution mass spectrometry

被引:41
作者
Yu, LL
Fassett, JD
Guthrie, WF
机构
[1] Natl Inst Stand & Technol, Div Analyt Chem, Gaithersburg, MD 20899 USA
[2] Natl Inst Stand & Technol, Stat Engn Div, Gaithersburg, MD 20899 USA
关键词
D O I
10.1021/ac011254l
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The detection limit is an important figure of merit for evaluating instrumentation and analytical methods. While the detection limit for techniques using linear calibration functions has been studied extensively, this fundamental metric has rarely been discussed for mass spectrometry that bases the calibration on the principle of isotope dilution.,We have developed a formulation for the detection-limit for isotope dilution mass spectrometry (IDMS) after a thorough analysis of the uncertainty of IDMS measurements. The new formulation describes the, IDMS detection limit as a function of the enrichment of the isotopic spike and the linear calibration detection limits measured at the masses for the isotope ratio measurement.
引用
收藏
页码:3887 / 3891
页数:5
相关论文
共 8 条
[1]  
Analytical Chemistry Division, 1978, SPECTROCHIM ACTA B, V33, P242
[2]  
[Anonymous], 1995, Guide to the expression of uncertainty in measurement (GUM)
[3]   NOMENCLATURE FOR THE PRESENTATION OF RESULTS OF CHEMICAL-ANALYSIS [J].
CURRIE, LA ;
SVEHLA, G .
PURE AND APPLIED CHEMISTRY, 1994, 66 (03) :595-608
[5]  
FASSETT JD, 1989, ANAL CHEM, V61, P643
[6]  
Heumann K.G, 1988, INORGANIC MASS SPECT
[7]   The optimal amount of isotopic spike solution for ultratrace analysis by isotope dilution mass spectrometry [J].
Hoelzl, R ;
Hoelzl, C ;
Kotz, L ;
Fabry, L .
ACCREDITATION AND QUALITY ASSURANCE, 1998, 3 (05) :185-188
[8]   Chromatographic isolation of molybdenum from human blood plasma and determination by inductively coupled plasma mass spectrometry with isotope dilution [J].
Luong, ET ;
Houk, RS ;
Serfass, RE .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1997, 12 (07) :703-708