Signal-to-noise ratio and spatial resolution in x-ray electronic imagers: Is the MTF a relevant parameter?

被引:63
作者
Moy, JP [1 ]
机构
[1] Zone Ind Ctr Alp, F-38430 Moirans, France
关键词
x-ray imaging; MTF; DQE; noise aliasing;
D O I
10.1118/1.598859
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In most imaging detectors, the modulation transfer function (MTF) is regarded as a good parameter to describe spatial resolution. This is undoubtedly valid far visual observation. However, the detectability of a detail is essentially a matter of signal-to-noise ratio, which is not accounted for by the MTF. In x-ray imaging, signal-to-noise ratio in the image is generally limited by incident photons statistics, often larger than readout noises. Therefore, the MTF of the detector applies to both signal and noise, and does not impair the image content. Contrast can easily be restored by image processing without altering the signal-to-noise ratio. However. a number of effects may alter very differently noise and signal: (i) If the MTF significantly extends beyond half the sampling frequency, the aliasing introduced by spatial sampling can severely enhance the noise and cancel the benefit of the good signal transfer. This is illustrated by synthetic images which simulate the response of imagers with different MTFs to the same test pattern in the presence of quantum noise. (ii) Parallax and blurring by the x-ray spot size or motion are shown to degrade the transfer properties of signal, but do not affect the quantum noise; they must be treated separately. Contrary to the x-ray converter MTF, parallax directly impacts the detective quantum efficiency (DQE). Finally, it is shown that only the detective quantum efficiency can reliably describe the spatial resolution of an x-ray imaging detector in the presence of noise, parallax and blurring. (C) 2000 American Association of Physicists in Medicine.
引用
收藏
页码:86 / 93
页数:8
相关论文
共 7 条
[1]   New CsI/a-Si 17" x 17" X-ray flat panel detector provides superior detectivity and immediate direct digital output for General Radiography systems [J].
Chaussat, C ;
Chabbal, J ;
Ducourant, T ;
Spinnler, V ;
Vieux, G ;
Neyret, R .
PHYSICS OF MEDICAL IMAGING, 1998, 3336 :45-56
[2]   Degradation of the detective quantum efficiency due to a non-unity detector fill factor [J].
Cunningham, IA .
PHYSICS OF MEDICAL IMAGING - MEDICAL IMAGING 1997, 1997, 3032 :22-31
[3]   A discussion on resolution and dynamic range of Se-TFT direct digital radiographic detector [J].
Lee, DL ;
Cheung, LK ;
Palecki, EF ;
Jeromin, LS .
PHYSICS OF MEDICAL IMAGING: MEDICAL IMAGING 1996, 1996, 2708 :511-522
[4]   RANDOM NOISE PRODUCED BY X-RAY FLUORESCENT SCREENS [J].
LUBBERTS, G .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA, 1968, 58 (11) :1475-&
[5]   RESULTS USING MEDIUM SUPPLEMENTED WITH EITHER FRESH OR FROZEN STORED SERUM IN HUMAN INVITRO FERTILIZATION [J].
SHAW, JM ;
HARRISON, KL ;
WILSON, LM ;
BREEN, TM ;
SHAW, G ;
CUMMINS, JM ;
HENNESSEY, JF .
JOURNAL OF IN VITRO FERTILIZATION AND EMBRYO TRANSFER, 1987, 4 (01) :5-9
[6]   CALCULATION OF MODULATION TRANSFER-FUNCTIONS OF X-RAY FLUORESCENT SCREENS [J].
SWANK, RK .
APPLIED OPTICS, 1973, 12 (08) :1865-1870
[7]   Screen design for flat-panel imagers in diagnostic radiology [J].
Trauernicht, DP ;
Yorkston, J .
PHYSICS OF MEDICAL IMAGING, 1998, 3336 :477-484