The art and science of forming packed analytical high-performance liquid chromatography columns

被引:77
作者
Kirkland, J. J. [1 ]
DeStefano, J. J. [1 ]
机构
[1] Adv Mat Technol Inc, St Petersburg 198103, Russia
关键词
high-performance liquid chromatography; packed analytical columns; slurry column packing methods; packing consolidation; packing particle compression; packed beds;
D O I
10.1016/j.chroma.2006.04.027
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Columns of packed particles still are the most popular devices for high-performance liquid chromatography (HPLC) separations because of their great utility, excellent performance and wide variety. However, the forming of packed beds for efficient, stable columns traditionally has been an art where the basics of how to form optimum beds generally was not well understood. The recent development of monolith rods was introduced in part to overcome the difficulty of producing stable beds of packing particles. However, these materials are less versatile than packed particle columns. Technology developments in recent years have produced a better understanding among those skilled in the practice of how to form optimized packed beds, and this has led to widely available, high-quality commercial columns. This presentation discusses the developments that led to the present state of column packing technology. Important steps in the packing of efficient, stable beds are described. The key step of selecting the best solvent for the slurry packing method is emphasized. Factors affecting the mechanical stability of packed columns also are discussed. The early art of packing columns now has evolved into a more scientific approach that allows the packing of good columns with a minimum of effort and time. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:50 / 57
页数:8
相关论文
共 23 条
[11]   Stability of silica-based, endcapped columns with pH 7 and 11 mobile phases for reversed-phase high-performance liquid chromatography [J].
Kirkland, JJ ;
Henderson, JW ;
DeStefano, JJ ;
vanStraten, MA ;
Claessens, HA .
JOURNAL OF CHROMATOGRAPHY A, 1997, 762 (1-2) :97-112
[14]   EFFECT OF COLUMN TO PARTICLE DIAMETER RATIO ON DISPERSION OF UNSORBED SOLUTES IN CHROMATOGRAPHY [J].
KNOX, JH ;
PARCHER, JF .
ANALYTICAL CHEMISTRY, 1969, 41 (12) :1599-&
[15]   KINETIC CONDITIONS FOR OPTIMUM SPEED AND RESOLUTION IN COLUMN CHROMATOGRAPHY [J].
KNOX, JH ;
SALEEM, M .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1969, 7 (10) :614-&
[16]  
NEUE UD, 1997, THEORY TECHNOLOGY PR, P94
[17]  
SCHOTT CD, 1971, MODERN PRACTICE LIQU, pCH8
[18]  
SNYDER LR, 1997, PRACTICAL HPLC METHO, pCH6
[19]  
SNYDER LR, 1979, INTRO MODERN LIQUID, pCH5
[20]  
SNYDER LR, 1997, PRACTICAL HPLC METHO, pCH5