Chloride selective potentiometric sensor based on a newly synthesized hydrogen bonding anion receptor

被引:57
作者
Gupta, Vinod K. [1 ]
Goyal, R. N. [1 ]
Sharma, Ram A. [1 ]
机构
[1] Indian Inst Technol Roorkee, Dept Chem, Roorkee 247667, Uttar Pradesh, India
关键词
Sensor; Potentiometry; PVC membrane; Chloride; Anion sensor; RECOGNITION; ELECTRODES; COEFFICIENTS;
D O I
10.1016/j.electacta.2009.02.068
中图分类号
O646 [电化学、电解、磁化学];
学科分类号
081704 ;
摘要
A new potentiometric chloride sensor based on the use of anion receptor 2-(1-H-imidazo [4,5-f][1,10]phenanthroline-2-yl)-6methoxyphenol (HIPM) in poly(vinyl chloride) (PVC) matrix is reported. Effect of various plasticizers: 2-nitrophenyloctylether (o-NPOE), di-n-butylphthalate (DBP), diethyl phthalate (DEP), dioctylpthalate (DOP), tri-n-butyl phosphates (TBP), chloronapthalene (CN) and cation excluder, cetryltrimethylammoniumbromide (CTAB) was studied. The best performance was obtained with a membrane composition of PVC: o-NPOE:HIPM: CTAB ratio (w/w, %) of 31:60:7:3. The sensor exhibits significantly enhanced selectivity towards chloride over the concentration range 5.0 x 10(-8) to 1.0 x 10(-1) M with a lower detection limit of 1.0 x 10(-8) M and a Nernstian slope of 59.8 mV decade(-1) of activity. Influence of the membrane composition and possible interfering ions was investigated on the response properties of the electrode. Fast and stable response, good reproducibility and long-term stability are demonstrated. The sensor shows response time of <10s; and can be used for about 3 months without any considerable divergence in their potential response. Selectivity coefficients determined with matched potential method (MPM) indicate high selectivity for chloride. The proposed electrode shows fairly good discrimination of chloride from anions. It was successfully applied to determination of chloride ion in packed drinking water. The electrode can be used in the pH range of 6.5-8.0 and mixtures containing up to 20% (v/v) non-aqueous content. It was used as an indicator electrode in potentiometric titration of chloride ion against silver nitrate. (C) 2009 Elsevier Ltd. All rights reserved.
引用
收藏
页码:4216 / 4222
页数:7
相关论文
共 40 条
[1]   Selectivity of potentiometric ion sensors [J].
Bakker, E ;
Pretsch, E ;
Bühlmann, P .
ANALYTICAL CHEMISTRY, 2000, 72 (06) :1127-1133
[2]   Carrier-based ion-selective electrodes and bulk optodes. 1. General characteristics [J].
Bakker, E ;
Buhlmann, P ;
Pretsch, E .
CHEMICAL REVIEWS, 1997, 97 (08) :3083-3132
[3]  
BANERJI SK, 1925, J SCI IND RES ACAD S, V180, P928
[4]   Transition metal and organometallic anion complexation agents [J].
Beer, PD ;
Hayes, EJ .
COORDINATION CHEMISTRY REVIEWS, 2003, 240 (1-2) :167-189
[5]  
Beer PD, 2001, ANGEW CHEM INT EDIT, V40, P486, DOI 10.1002/1521-3773(20010202)40:3<486::AID-ANIE486>3.0.CO
[6]  
2-P
[7]   Abiotic guanidinium containing receptors for anionic species [J].
Best, MD ;
Tobey, SL ;
Anslyn, EV .
COORDINATION CHEMISTRY REVIEWS, 2003, 240 (1-2) :3-15
[8]  
BINAS FG, 2005, IEEE, P43
[9]   Amide based receptors for anions [J].
Bondy, CR ;
Loeb, SJ .
COORDINATION CHEMISTRY REVIEWS, 2003, 240 (1-2) :77-99
[10]   Confirmation of a 'cleft-mode' of binding in a 2,5-diamidopyrrole anion receptor in the solid state [J].
Camiolo, S ;
Gale, PA ;
Hursthouse, MB ;
Light, ME .
TETRAHEDRON LETTERS, 2002, 43 (39) :6995-6996