Experimental investigation of the steady flow downstream of the St. Jude bileaflet heart valve: A comparison between laser Doppler velocimetry and particle image velocimetry techniques

被引:35
作者
Browne, P
Ramuzat, A
Saxena, R
Yoganathan, AP
机构
[1] Georgia Inst Technol, Sch Biomed Engn, Cardiovasc Fluid Mech Lab, Atlanta, GA 30332 USA
[2] Von Karman Inst Fluid Dynam, Appl Fluid Dynam Dept, Rhode St Genese, Belgium
关键词
laser doppler velocimetry; particle image velocimetry; velocity field; maximum turbulent stress; prosthetic heart valve;
D O I
10.1114/1.252
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This study investigates turbulent flow, based on high Reynolds number, downstream of a prosthetic heart valve using both laser Doppler velocimetry (LDV) and particle image velocimetry (PIV). Until now, LDV has been the more commonly used tool in investigating the flow characteristics associated with mechanical heart valves. The LDV technique allows point by point velocity measurements and provides enough statistical information to quantify turbulent structure. The main drawback of this technique is the time consuming nature of the data acquisition process in order to assess an entire flow field area. Another technique now used in fluid dynamics studies is the PIV measurement technique. This technique allows spatial and temporal measurement of the entire flow field. Using this technique, the instantaneous and average velocity how fields can be investigated for different positions. This paper presents a comparison of PIV two-dimensional measurements to LDV measurements, performed under steady flow conditions, for a measurement plane parallel to the leaflets of a St. Jude Medical (SJM) bileaflet valve. Comparisons of mean velocity obtained by the two techniques are in good agreement except for where there is instability in the flow. For second moment quantities the comparisons were less agreeable. This suggests that the PIV technique has sufficient temporal and spatial resolution to estimate mean velocity depending on the degree of instability in the flow and also provides sufficient images needed to duplicate mean flow but not for higher moment turbulence quantities such as maximum turbulent shear stress, (C) 2000 Biomedical Engineering Society. [S0090-6964(00)01001-8].
引用
收藏
页码:39 / 47
页数:9
相关论文
共 13 条
  • [1] Bruss K H, 1983, Life Support Syst, V1, P3
  • [2] Identification of peak stresses in cardiac prostheses - A comparison of two-dimensional versus three-dimensional principal stress analyses
    Fontaine, AA
    Ellis, JT
    Healy, TM
    Hopmeyer, J
    Yoganathan, AP
    [J]. ASAIO JOURNAL, 1996, 42 (03) : 154 - 163
  • [3] An automated method for analysis and visualization of laser Doppler velocimetry data
    Healy, TM
    Ellis, JT
    Fontaine, AA
    Jarrett, CA
    Yoganathan, AP
    [J]. ANNALS OF BIOMEDICAL ENGINEERING, 1997, 25 (02) : 335 - 343
  • [4] KARATEKIN O, IN PRESS 9 INT S APP
  • [5] Steady flow velocity field and turbulent stress mappings downstream of a porcine bioprosthetic aortic valve in vitro
    Lim, WL
    Chew, YT
    Chew, TC
    Low, HT
    [J]. ANNALS OF BIOMEDICAL ENGINEERING, 1997, 25 (01) : 86 - 95
  • [6] Steady flow dynamics of prosthetic aortic heart valves: a comparative evaluation with PIV techniques
    Lim, WL
    Chew, YT
    Chew, TC
    Low, HT
    [J]. JOURNAL OF BIOMECHANICS, 1998, 31 (05) : 411 - 421
  • [7] LOURENCO LM, 1982, THESIS U LIBRE BRUXE
  • [8] MUNSON BR, 1998, FUNDAMENTALS FLUID M, P480
  • [9] WOO YR, 1985, LIFE SUPPORT SYST, V3, P283
  • [10] INVITRO VELOCITY-MEASUREMENTS IN THE VICINITY OF AORTIC PROSTHESES
    YOGANATHAN, AP
    CORCORAN, WH
    HARRISON, EC
    [J]. JOURNAL OF BIOMECHANICS, 1979, 12 (02) : 135 - 152