Simulating oblique incident irradiation using the BEAMnrc Monte Carlo code

被引:5
作者
Downes, P. [1 ,2 ]
Spezi, E. [2 ,3 ]
机构
[1] Cardiff Univ, Cardiff Sch Comp Sci, Cardiff CF24 3AA, S Glam, Wales
[2] Velindre Canc Ctr, Dept Med Phys, Cardiff CF14 2TL, S Glam, Wales
[3] Cardiff Univ, Sch Engn, Inst Med Engn & Med Phys, Cardiff CF24 3AA, S Glam, Wales
基金
英国工程与自然科学研究理事会;
关键词
D O I
10.1088/0031-9155/54/7/N02
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A new source for the simulation of oblique incident irradiation has been developed for the BEAMnrc Monte Carlo code. In this work, we describe a method for the simulation of any component that is rotated at some angle relative to the central axis of the modelled radiation unit. The performance of the new BEAMnrc source was validated against experimental measurements. The comparison with ion chamber data showed very good agreement between experiments and calculation for a number of oblique irradiation angles ranging from 0 degrees to 30 degrees. The routine was also cross-validated, in geometrically equivalent conditions, against a different radiation source available in the DOSXYZnrc code. The test showed excellent consistency between the two routines. The new radiation source can be particularly useful for the Monte Carlo simulation of radiation units in which the radiation beam is tilted with respect to the unit's central axis. To highlight this, a modern cone-beam CT unit is modelled using this new source and validated against measurement.
引用
收藏
页码:N93 / N100
页数:8
相关论文
共 10 条
[1]   An evaluation of epoxy resin phantom materials for megavoltage photon dosimetry [J].
Allahverdi, M ;
Nisbet, A ;
Thwaites, DI .
PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (05) :1125-1132
[2]   A SOLID WATER PHANTOM MATERIAL FOR RADIOTHERAPY X-RAY AND GAMMA-RAY BEAM CALIBRATIONS [J].
CONSTANTINOU, C ;
ATTIX, FH ;
PALIWAL, BR .
MEDICAL PHYSICS, 1982, 9 (03) :436-441
[3]  
Foley J.D., 1996, Computer graphics: Principles and practice
[4]   Large efficiency improvements in BEAMnrc using directional bremsstrahlung splitting [J].
Kawrakow, I ;
Rogers, DWO ;
Walters, BRB .
MEDICAL PHYSICS, 2004, 31 (10) :2883-2898
[5]   Accurate condensed history Monte Carlo simulation of electron transport. I. EGSnrc, the new EGS4 version [J].
Kawrakow, I .
MEDICAL PHYSICS, 2000, 27 (03) :485-498
[6]   Fifty years of Monte Carlo simulations for medical physics [J].
Rogers, D. W. O. .
PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (13) :R287-R301
[7]  
Rogers D W O, 2007, PIRS0509AREVK NRCC
[8]   BEAM - A MONTE-CARLO CODE TO SIMULATE RADIOTHERAPY TREATMENT UNITS [J].
ROGERS, DWO ;
FADDEGON, BA ;
DING, GX ;
MA, CM ;
WE, J ;
MACKIE, TR .
MEDICAL PHYSICS, 1995, 22 (05) :503-524
[9]  
WALTERS B, 2007, PIRS794REVB NRCC
[10]  
WALTERS B, 2008, COMMUNICATION