The potential of gas chromatography with microwave-induced plasma atomic emission detection (GC-MIP-AED) as a complementary analytical technique in environmental screening analysis of aqueous samples

被引:14
作者
Frischenschlager, H [1 ]
Mittermayr, C [1 ]
Peck, M [1 ]
Rosenberg, E [1 ]
Grasserbauer, M [1 ]
机构
[1] VIENNA UNIV TECHNOL,INST ANALYT CHEM,A-1060 VIENNA,AUSTRIA
来源
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY | 1997年 / 359卷 / 03期
基金
奥地利科学基金会;
关键词
D O I
10.1007/s002160050563
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
With GC-MIP-AED as a complementary analytical technique to GC-MS, the reliable identification of many organic pollutants in water samples of the Nitra river has been made possible due to the additional information on the elemental composition of organic pollutants it provides. Some compounds, which have been identified by GC-MS in the scan mode after a MS-Library search, can be confirmed by their element responses. Even where the identification of substances cannot be confirmed by their empirical formulas calculated by GC-MIP-AED due to insufficient peak intensities, integration problems or chromatographic interferences, it can be facilitated by the presence of one or more heteroatoms. Quantitative data of selected organic pollutants from GC-MTP-AED screening analysis after compound independent calibration can be obtained more easily than, and in good agreement with, quantitative results provided by target analysis (e.g. GCMS in the SIM-mode). New self-developed macros are helpful tools for acquiring information on the elemental composition and concentration of organic pollutants analyzed by GC-MIP-AED.
引用
收藏
页码:213 / 221
页数:9
相关论文
共 34 条
[1]   CAVITY FOR MICROWAVE-INDUCED PLASMAS OPERATED IN HELIUM AND ARGON AT ATMOSPHERIC-PRESSURE [J].
BEENAKKER, CIM .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1976, 31 (8-9) :483-486
[2]   CONSIDERATIONS IN THE DESIGN OF A MICROWAVE INDUCED PLASMA UTILIZING THE TM010 CAVITY FOR OPTICAL-EMISSION SPECTROSCOPY [J].
BOLLOKAMARA, A ;
CODDING, EG .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1981, 36 (10) :973-982
[3]  
CASABIANCA H, 1992, ANALUSIS S7, V20, P9
[4]  
DAVID F, 1993, LC GC-MAG SEP SCI, V11, P282
[5]  
DAVID F, 1991, HRC-J HIGH RES CHROM, V14, P554
[6]  
DAVID F, 1992, SPECTRA 2000, V167, P31
[7]   ANALYTICAL STRATEGY BY COUPLING HEADSPACE GAS-CHROMATOGRAPHY, ATOMIC-EMISSION SPECTROMETRIC DETECTION AND MASS-SPECTROMETRY APPLICATION TO SULFUR-COMPOUNDS FROM GARLIC [J].
DERUAZ, D ;
SOUSSANMARCHAL, F ;
JOSEPH, I ;
DESAGE, M ;
BANNIER, A ;
BRAZIER, JL .
JOURNAL OF CHROMATOGRAPHY A, 1994, 677 (02) :345-354
[8]   DETERMINATION OF RATIO FORMULAS FOR ORGANIC-COMPOUNDS USING A MICROWAVE INDUCED PLASMA [J].
DINGJAN, HA ;
DEJONG, HJ .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1983, 38 (5-6) :777-781
[9]   ELEMENT-SELECTIVE DETECTION OF PESTICIDES BY GAS-CHROMATOGRAPHY ATOMIC-EMISSION DETECTION AND SOLID-PHASE MICROEXTRACTION [J].
EISERT, R ;
LEVSEN, K ;
WUNSCH, G .
JOURNAL OF CHROMATOGRAPHY A, 1994, 683 (01) :175-183
[10]   Improved screening analysis of organic pollutants in river water samples by gas chromatography with atomic emission detection (GC-MIP-AED) [J].
Frischenschlager, H ;
Peck, ML ;
Mittermayr, C ;
Rosenberg, E ;
Grasserbauer, M .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1997, 357 (08) :1133-1141