Reproducibility and sensitivity of scanning laser Doppler flowmetry during graded changes in Po-2

被引:78
作者
Strenn, K
Menapace, R
Rainer, G
Findl, O
Wolzt, M
Schmetterer, L
机构
[1] UNIV VIENNA, DEPT CLIN PHARMACOL, A-1090 VIENNA, AUSTRIA
[2] UNIV VIENNA, DEPT OPHTHALMOL B, A-1090 VIENNA, AUSTRIA
[3] UNIV VIENNA, INST MED PHYS, A-1090 VIENNA, AUSTRIA
关键词
D O I
10.1136/bjo.81.5.360
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Aims/background - Recently a commercially available scanning laser Doppler flowmeter has been produced, which provides two dimensional maps of the retinal perfusion. The aim of the present study was to investigate the reproducibility and the sensitivity of these measurements. Methods - 16 healthy subjects were randomised to inhale different gas mixtures of oxygen and nitrogen in a double blind crossover study. The following gas mixtures of oxygen and nitrogen were administered: 100% oxygen + 0% nitrogen, 80% oxygen + 20% nitrogen, 60% oxygen + 40% nitrogen, 40% oxygen + 60% nitrogen, 30% oxygen + 70% nitrogen, 20% oxygen + 80% nitrogen, 15% oxygen + 85% nitrogen, and 10% oxygen + 90% nitrogen. Retinal haemodynamic variables and systemic haemodynamics were measured during all inhalation periods. Recordings under resting conditions were performed three times to calculate intraclass coefficients. Results - In two subjects we did not obtain technically adequate results. A dose dependent change in retinal blood flow during graded oxygen breathing was observed (p < 0.001). During 100% oxygen breathing blood how decrease was between 29% and 33%, whereas blood flow increase was between 28% and 33% during inhalation of 10% oxygen + 90% nitrogen. Conclusions - Scanning laser Doppler flowmetry has an acceptable reproducibility and is appropriate for description of the effect of graded changes in PO2 on retinal haemodynamics. The main problems with the system are the large zero offset, the fixation during retinal scanning, and the neglect of blood flow changes during the cardiac cycle.
引用
收藏
页码:360 / 364
页数:5
相关论文
共 25 条
[1]  
[Anonymous], [No title captured]
[2]   MACULAR CAPILLARY PARTICLE VELOCITIES - A BLUE FIELD AND SCANNING LASER COMPARISON [J].
AREND, O ;
HARRIS, A ;
SPONSEL, WE ;
REMKY, A ;
REIM, M ;
WOLF, S .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 1995, 233 (04) :244-249
[3]  
FALLON TJ, 1985, OPHTHALMOLOGY, V92, P701
[4]   CHOROIDAL AND CILIARY BODY BLOOD-FLOW ANALYSIS - APPLICATION OF LASER DOPPLER FLOWMETRY IN EXPERIMENTAL-ANIMALS [J].
GHEREZGHIHER, T ;
OKUBO, H ;
KOSS, MC .
EXPERIMENTAL EYE RESEARCH, 1991, 53 (02) :151-156
[5]   OPTIC-NERVE HEAD BLOOD-FLOW USING A LASER-DOPPLER VELOCIMETER AND HEMORHEOLOGY IN PRIMARY OPEN-ANGLE GLAUCOMA AND NORMAL-PRESSURE GLAUCOMA [J].
HAMARD, P ;
HAMARD, H ;
DUFAUX, J ;
QUESNOT, S .
BRITISH JOURNAL OF OPHTHALMOLOGY, 1994, 78 (06) :449-453
[6]  
Harris A, 1995, J Glaucoma, V4, P281
[7]   CLINICAL BIOSTATISTICS .54. THE BIOSTATISTICS OF CONCORDANCE [J].
KRAMER, MS ;
FEINSTEIN, AR .
CLINICAL PHARMACOLOGY & THERAPEUTICS, 1981, 29 (01) :111-123
[8]  
LIETZ A, 1996, INVEST OPHTHALMOL S, V37, P462
[9]  
LUNDMARK P, 1996, INVEST OPHTHALMOL S, V37, P265
[10]   2-DIMENSIONAL MAPPING OF RETINAL AND PAPILLARY MICROCIRCULATION BY SCANNING LASER-DOPPLER FLOWMETRY [J].
MICHELSON, G ;
GROH, M ;
LANGHANS, M ;
SCHMAUSS, B .
KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 1995, 207 (03) :180-190