Continuous intraoperative noninvasive cardiac output monitoring using a new thoracic bioimpedance device

被引:23
作者
Thangathurai, D [1 ]
Charbonnet, C [1 ]
Roessler, P [1 ]
Wo, CCJ [1 ]
Mikhail, M [1 ]
Yoshida, R [1 ]
Shoemaker, WC [1 ]
机构
[1] UNIV SO CALIF,SCH MED,DEPT ANESTHESIOL,LOS ANGELES,CA
关键词
measurement of cardiac output; thoracic bioimpedance; noninvasive cardiac output measurement; thermodilution cardiac output; continuous hemodynamic monitoring; high-risk surgery; pulmonary artery catheterization; ELECTRICAL BIOIMPEDANCE; IMPEDANCE CARDIOGRAPHY; THERMODILUTION;
D O I
10.1016/S1053-0770(97)90052-5
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Objectives: To compare a new noninvasive bioimpedance device with the standard thermodilution method during the intraoperative period in high-risk patients undergoing oncological surgery. Design: Prospectively collected data with retrospective analysis. Setting: The study was undertaken at a university hospital, single institution. Participants: Twenty-three selected adults undergoing extensive, ablative oncological surgery. Interventions: Simultaneous measurements of cardiac output by a new bioimpedance method and the standard thermodilution method during the intraoperative and immediate postoperative periods. Measurements and Main Results: The correlation coefficient between the two methods was r = 0.89, p < 0.001. Bias and precision analysis between the two techniques showed a mean bias of 0.1 L/min and SD of the bias (precision) of 1.0 L/min (95% level of agreement +2.1 L/min to -1.9 L/min). After software enhancement, data from the last 11 monitored patients showed improved correlation between the two methods; r = 0.93, mean bias -0.1 L/min, and precision 0.8 L/min,Electrical and motion induced interference only transiently impaired the performance of the new impedance method. Conclusion: This new impedance device is a safe, reliable, clinically acceptable alternative to the invasive thermodilution method in the operating room environment. Copyright (C) 1997 by W.B. Saunders Company.
引用
收藏
页码:440 / 444
页数:5
相关论文
共 20 条
  • [1] COMPARISON OF MEASUREMENTS OF CARDIAC-OUTPUT BY BIOIMPEDANCE AND THERMODILUTION IN SEVERELY ILL SURGICAL PATIENTS
    APPEL, PL
    KRAM, HB
    MACKABEE, J
    FLEMING, AW
    SHOEMAKER, WC
    [J]. CRITICAL CARE MEDICINE, 1986, 14 (11) : 933 - 935
  • [2] CONTINUOUS NONINVASIVE REAL-TIME MONITORING OF STROKE VOLUME AND CARDIAC-OUTPUT BY THORACIC ELECTRICAL BIOIMPEDANCE
    BERNSTEIN, DP
    [J]. CRITICAL CARE MEDICINE, 1986, 14 (10) : 898 - 901
  • [3] RELATIONSHIP BETWEEN SUPRANORMAL CIRCULATORY VALUES, TIME DELAYS, AND OUTCOME IN SEVERELY TRAUMATIZED PATIENTS
    BISHOP, MH
    SHOEMAKER, WC
    APPEL, PL
    WO, CJ
    ZWICK, C
    KRAM, HB
    MEADE, P
    KENNEDY, F
    FLEMING, AW
    [J]. CRITICAL CARE MEDICINE, 1993, 21 (01) : 56 - 63
  • [4] BLAND JM, 1986, LANCET, V1, P407
  • [5] MEASUREMENT OF CARDIAC-OUTPUT BY IMPEDANCE CARDIOGRAPHY UNDER VARIOUS CONDITIONS
    BOER, P
    ROOS, JC
    GEYSKES, GG
    DORHOUTMEES, EJ
    [J]. AMERICAN JOURNAL OF PHYSIOLOGY, 1979, 237 (04): : H491 - H496
  • [6] DETERMINATION OF CARDIAC-OUTPUT DURING POSITIVE ENDEXPIRATORY PRESSURE - NONINVASIVE ELECTRICAL BIOIMPEDANCE COMPARED WITH STANDARD THERMODILUTION
    CASTOR, G
    MOLTER, G
    HELMS, J
    NIEDERMARK, I
    ALTMAYER, P
    [J]. CRITICAL CARE MEDICINE, 1990, 18 (05) : 544 - 546
  • [7] CARDIAC-OUTPUT MEASUREMENT IN CRITICAL CARE PATIENTS - THORACIC ELECTRICAL BIOIMPEDANCE VERSUS THERMODILUTION
    CLANCY, TV
    NORMAN, K
    REYNOLDS, R
    COVINGTON, D
    MAXWELL, JG
    [J]. JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 1991, 31 (08): : 1116 - 1121
  • [8] PREDICTORS OF BETWEEN-METHOD DIFFERENCES IN CARDIAC-OUTPUT MEASUREMENT USING THORACIC ELECTRICAL BIOIMPEDANCE AND THERMODILUTION
    DOERING, L
    LUM, E
    DRACUP, K
    FRIEDMAN, A
    [J]. CRITICAL CARE MEDICINE, 1995, 23 (10) : 1667 - 1673
  • [9] COMPARISON OF TRANS-THORACIC ELECTRICAL-IMPEDANCE AND THERMODILUTION METHODS FOR MEASURING CARDIAC-OUTPUT
    DONOVAN, KD
    DOBB, GJ
    WOODS, WPD
    HOCKINGS, BE
    [J]. CRITICAL CARE MEDICINE, 1986, 14 (12) : 1038 - 1044
  • [10] MEASUREMENT OF CARDIAC-OUTPUT IN PREGNANCY BY THORACIC ELECTRICAL BIOIMPEDANCE AND THERMODILUTION - A PRELIMINARY-REPORT
    MASAKI, DI
    GREENSPOON, JS
    OUZOUNIAN, JG
    [J]. AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1989, 161 (03) : 680 - 684