Interdigitated microelectrode (IME) impedance sensor for the detection of viable Salmonella typhimurium

被引:244
作者
Yang, LJ [1 ]
Li, YB
Griffis, CL
Johnson, MG
机构
[1] Univ Arkansas, Dept Biol & Agr Engn, Fayetteville, AR 72701 USA
[2] Univ Arkansas, Ctr Excellence Poultry Sci, Fayetteville, AR 72701 USA
[3] Univ Arkansas, Dept Food Sci, Fayetteville, AR 72701 USA
关键词
interdigitated microelectrode; impedance sensor; bacteria detection; Salmonella typhimurium; biosensing;
D O I
10.1016/j.bios.2003.10.009
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Interdigitated microelectrodes (IMEs) were used as impedance sensors for rapid detection of viable Salmonella typhimurium in a selective medium and milk samples. The impedance growth curves, impedance against bacterial growth time, were recorded at four frequencies (10 Hz, 100 Hz, I kHz, and 10 kHz) during the growth of S. typhimurium. The impedance did not change until the cell number reached 10(5) - 10(6) CFU ml(-1). The greatest change in impedance was observed at 10 Hz. To better understand the mechanism of the IME impedance sensor, an equivalent electrical circuit, consisting of double layer capacitors, a dielectric capacitor, and a medium resistor, was introduced and used for interpreting the change in impedance during bacterial growth. Bacterial attachment to the electrode surface was observed with scanning electron microscopy, and it had effect on the impedance measurement. The detection time, t(D), defined as the time for the impedance to start change, was obtained from the impedance growth curve at 10 Hz and had a linear relationship with the logarithmic value of the initial cell number of S. typhimurium in the medium and milk samples. The regression equations for the cell numbers between 4.8 and 5.4 x 10(5) CFU ml(-1) were t(D) = -1.38 log N + 10.18 with R-2 = 0.99 in the pure medium and t(D) = -1.54 log N + 11.33 with R-2 = 0.98 in milk samples, respectively. The detection times for 4.8 and 5.4 x 10(5) CFU ml(-1) initial cell numbers were 9.3 and 2.2 h, respectively, and the detection limit could be as low as I cell in a sample. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:1139 / 1147
页数:9
相关论文
共 35 条
[1]   Pilot workload: Contemporary issues [J].
Andre, AD ;
Hancock, PA .
INTERNATIONAL JOURNAL OF AVIATION PSYCHOLOGY, 1995, 5 (01) :1-4
[2]  
[Anonymous], 1996, OFFICIAL METHODS ANA, V16th
[3]   RAPID AND AUTOMATED DETECTION OF SALMONELLA BY ELECTRICAL MEASUREMENTS [J].
EASTER, MC ;
GIBSON, DM .
JOURNAL OF HYGIENE, 1985, 94 (03) :245-262
[4]  
Eden R., 1984, IMPEDANCE MICROBIOLO
[5]   Monitoring of cellular behaviour by impedance measurements on interdigitated electrode structures [J].
Ehret, R ;
Baumann, W ;
Brischwein, M ;
Schwinde, A ;
Stegbauer, K ;
Wolf, B .
BIOSENSORS & BIOELECTRONICS, 1997, 12 (01) :29-41
[6]   On-line control of cellular adhesion with impedance measurements using interdigitated electrode structures [J].
Ehret, R ;
Baumann, W ;
Brischwein, M ;
Schwinde, A ;
Wolf, B .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1998, 36 (03) :365-370
[7]   Impedance microbiology: quantification of bacterial content in milk by means of capacitance growth curves [J].
Felice, CJ ;
Madrid, RE ;
Olivera, JM ;
Rotger, VI ;
Valentinuzzi, ME .
JOURNAL OF MICROBIOLOGICAL METHODS, 1999, 35 (01) :37-42
[8]   IMPEDANCE BACTERIOMETRY - MEDIUM AND INTERFACE CONTRIBUTIONS DURING BACTERIAL-GROWTH [J].
FELICE, CJ ;
VALENTINUZZI, ME ;
VERCELLONE, MI ;
MADRID, RE .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1992, 39 (12) :1310-1313
[9]   Medium and interface components in impedance microbiology [J].
Felice, CJ ;
Valentinuzzi, ME .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1999, 46 (12) :1483-1487
[10]  
Futschik K., 1988, INT M CHEM ENG BIOT, V88, P3