Urea rebound and delivered Kt/V determination with a continuous urea sensor

被引:25
作者
Garred, LJ
Canaud, B
Bosc, JY
Tetta, C
机构
[1] LAPEYRONIE HOSP, SERV NEPHROL, MONTPELLIER, FRANCE
[2] BELLCO SPA, MIRANDOLA, ITALY
关键词
haemodialysis; Kt/V; mathematical modelling; rebound; urea kinetic modelling; urea sensor;
D O I
10.1093/ndt/12.3.535
中图分类号
R3 [基础医学]; R4 [临床医学];
学科分类号
1001 ; 1002 ; 100602 ;
摘要
Background. The recent introduction of urea sensors for dialysis monitoring has made possible new approaches to urea kinetic modelling. In this study we show how the equilibrated postdialysis urea concentration (C-eq) and Kt/V corrected for double-pool urea kinetics (Kt/V-dp) can be accurately determined using an on-line sensor providing a continuous measure of blood water urea. A modification of the Smye constant volume double-pool theory led to the following equations for C-eq and Kt/V-dp C-eq = C-pre exp(S(ex)t) [GRAPHICS] where C-pre is the blood concentration measured at the start of dialysis, t is the length of the dialysis session (in min) and S-ex is the constant slope of the blood urea logarithm concentration decline following development of the intercompartmental urea concentration gradient in the first 30-60 min of dialysis. Methods, These equations were tested in 11 patients undergoing 165-240 min of paired filtration dialysis with continuous monitoring of blood urea concentration. C-pre was determined as the plateau concentration during a preliminary period of 15-20 min of slow isolated ultrafiltration. S-ex was accurately determined from linear regression applied to the urea sensor data from the 80-min point to the end of dialysis. Results. C-eq and Kt/V-dp determined from the above equations compared closely to values determined from 25-40 min of urea rebound monitoring with the urea sensor: 10.6 +/- 3.0 versus 10.8 +/- 2.7 mmol/l (mean +/- SD) for C-eq and 1.21 +/- 0.24 versus 1.18 +/- 0.20 for Kt/V-dp, compared to single-pool values of Kt/V = 1.34 +/- 0.23. Conclusion, This technique may be readily programmed into on-line urea monitors to provide current and extrapolated values of C-eq and Kt/V-dp from about the first hour of dialysis.
引用
收藏
页码:535 / 542
页数:8
相关论文
共 15 条
[11]  
Santoro A, 1996, NEPHROL DIAL TRANSPL, V11, P1084
[12]   A mathematical analysis of a two-compartment model of urea kinetics [J].
Smye, SW ;
Will, EJ .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (12) :2005-2014
[13]   ESTIMATION OF TREATMENT DOSE IN HIGH-EFFICIENCY HEMODIALYSIS [J].
SMYE, SW ;
DUNDERDALE, E ;
BROWNRIDGE, G ;
WILL, E .
NEPHRON, 1994, 67 (01) :24-29
[14]   PEDIATRIC HEMODIALYSIS - ESTIMATION OF TREATMENT EFFICIENCY IN THE PRESENCE OF UREA REBOUND [J].
SMYE, SW ;
EVANS, JHC ;
WILL, E ;
BROCKLEBANK, JT .
CLINICAL PHYSICS AND PHYSIOLOGICAL MEASUREMENT, 1992, 13 (01) :51-62
[15]  