Validation and use of the Finometer™ for blood pressure measurement in normal, hypertensive and pre-eclamptic pregnancy

被引:30
作者
Elvan-Taspinar, A
Uiterkamp, LAMS
Sikkema, JM
Bots, ML
Koomans, HA
Bruinse, HW
Franx, A
机构
[1] Univ Utrecht, Med Ctr, Dept Perinatol & Gynaecol, NL-3508 AB Utrecht, Netherlands
[2] Univ Utrecht, Med Ctr, Julius Ctr Hlth Sci & Primary Care, NL-3508 AB Utrecht, Netherlands
[3] Univ Utrecht, Med Ctr, Dept Nephrol, NL-3508 AB Utrecht, Netherlands
关键词
finger arterial pressure; automated blood pressure measurement; validation study; pregnancy; hypertension; pre-eclampsia;
D O I
10.1097/01.hjh.0000098138.70956.68
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objective Although a large variety of automated blood pressure devices are available, only some have been validated for use in clinical practice. The British Hypertension Society (BHS) recommends separate validation of automated devices in special subgroups, e.g. the elderly and pregnant women. The aim of this study was to compare the Finometer(TM) (FM) and the earlier validated SpaceLabs 90207(TM) (SL) with standard auscultatory blood pressure measurements in normal, pre-eclamptic and hypertensive pregnancy, following the guidelines of the BHS and the Association for the Advancement of Medical Instrumentation (AAMI). Methods The total study group consisted of 123 pregnant women, of whom were 54 normotensive, 31 pre-eclamptic and 38 hypertensive. Automated readings with the FM and SL were compared with auscultatory blood pressure measurements. Bland-Altman plots, BHS grades, mean pressure differences and 95% limits of agreement were used for analysis. Results Bland-Altman plots showed a wide scatter of the pressure differences between auscultatory and automated measurements. FM achieved BHS grades C/D, C/B, D/D and D/D in the total, normotensive, pre-eclamptic andhypertensive group, respectively. The AAMI criteria were only met for diastolic blood pressure in the normotensive group. For SL almost identical BHS grades and 95% limits of agreement as compared to our earlier study were found. Conclusions The accuracy and precision of the Finometer(TM) are not sufficient for determination of absolute blood pressure levels in individual pregnant women. Our present findings on the SpaceLabs 90207(TM) reconfirm our earlier results. (C) 2003 Lippincott Williams Wilkins.
引用
收藏
页码:2053 / 2060
页数:8
相关论文
共 30 条
[1]  
Baumgart, 1998, Blood Press Monit, V3, P303
[2]   COMPARING 2 METHODS OF CLINICAL MEASUREMENT - A PERSONAL HISTORY [J].
BLAND, JM ;
ALTMAN, DG .
INTERNATIONAL JOURNAL OF EPIDEMIOLOGY, 1995, 24 :S7-S14
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   THE RELIABILITY OF NONINVASIVE CONTINUOUS FINGER BLOOD-PRESSURE MEASUREMENT IN PATIENTS WITH BOTH HYPERTENSION AND VASCULAR-DISEASE [J].
BOS, WJW ;
IMHOLZ, BPM ;
VANGOUDOEVER, J ;
WESSELING, KH ;
VANMONTFRANS, GA .
AMERICAN JOURNAL OF HYPERTENSION, 1992, 5 (08) :529-535
[5]   Reconstruction of brachial artery pressure from noninvasive finger pressure measurements [J].
Bos, WJW ;
vanGoudoever, J ;
vanMontfrans, GA ;
vandenMeiracker, AH ;
Wesseling, KH .
CIRCULATION, 1996, 94 (08) :1870-1875
[6]   THE CLASSIFICATION AND DEFINITION OF THE HYPERTENSIVE DISORDERS OF PREGNANCY [J].
DAVEY, DA ;
MACGILLIVRAY, I .
AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1988, 158 (04) :892-898
[7]  
FRANX A, 1994, AM J OBSTET GYNECOL, V170, P702, DOI 10.1016/S0002-9378(94)70260-8
[8]   Comparison of an auscultatory versus an oscillometric ambulatory blood pressure monitor in normotensive, hypertensive, and preeclamptic pregnancy [J].
Franx, A ;
vanderPost, JAM ;
vanMontfrans, GA ;
Bruinse, HW .
HYPERTENSION IN PREGNANCY, 1997, 16 (02) :187-202
[9]   VALIDATION OF AUTOMATED BLOOD-PRESSURE RECORDING IN PREGNANCY [J].
FRANX, A ;
VANDERPOST, JAM ;
ELFERING, IM ;
VEERMAN, DP ;
MERKUS, HMWM ;
BOER, K ;
VANMONTFRANS, GA .
BRITISH JOURNAL OF OBSTETRICS AND GYNAECOLOGY, 1994, 101 (01) :66-69
[10]   Sympathetic nerve discharge in normal pregnancy and pregnancy-induced hypertension [J].
Greenwood, JP ;
Stoker, JB ;
Walker, JJ ;
Mary, DASG .
JOURNAL OF HYPERTENSION, 1998, 16 (05) :617-624