Relation of stent design and stent surface material to subsequent in-stent intimal hyperplasia in coronary arteries determined by intravascular ultrasound

被引:57
作者
Hoffmann, R
Mintz, GS
Haager, PK
Bozoglu, T
Grube, E
Gross, M
Beythien, C
Mudra, H
vom Dahl, J
Hanrath, P
机构
[1] Rhein Westfal TH Aachen, Med Clin 1, D-52057 Aachen, Germany
[2] Cardiovasc Res Fdn, New York, NY USA
[3] Heart Ctr Siegburg, Dept Cardiol, Siegburg, Germany
[4] Univ Clin Charite, Dept Cardiol, Berlin, Germany
[5] Univ Hamburg, Dept Cardiol, Hamburg, Germany
[6] Krankenhaus Neuperlach, Dept Cardiol, Munich, Germany
关键词
D O I
10.1016/S0002-9149(02)02347-0
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A variety of different stent designs and coatings have become available. This study sought to determine the impact of stent design and gold-coating of stents on intimal hyperplasia (IH) in human atherosclerotic coronary arteries in relation to known predictors of restenosis. Angiographic and intravascular ultrasound (IVUS) studies were performed at 6-month follow-up on 311 native coronary lesions of 311 patients treated with 99 Multi-Link stents, 74 InFlow steel stents, 73 InFlow gold-coated stents, 41 Palmaz-Schatz stents, 12 NIR steel stents, and 12 gold-coated NIR Royal stents. Lumen and stent cross-sectional area (CSA) were measured at 1-mm axial increments. Mean IH CSA (stent CSA - lumen CSA) and mean IH thickness were calculated and averaged over the total stent length. IVUS demonstrated different levels of IH for the 6 stents. Mean IH thickness ranged from 0.20 +/- 0.13 mm for Multi-Link stents to 0.43 +/- 0.14 mm for InFlow goal-coated stents (p <0.001). Multivariate analysis proved non-Multi-Link stent design (odds ratio 3.45, 95% confidence intervals 1.13 to 11.11, p <0.034) and gold coating (odds ratio 3.78, 95% confidence intervals 1.88 to 7.54, p <0.001) to be the only independent predictors of IH thickness >0.3 mm. In conclusion, stent design and surface material have an important impact on the IH response to stents implanted in human coronary arteries. However, the differences in IH thickness between the analyzed stents were relatively small compared with the absolute lumen dimensions. (C) 2002 by Excerpts Medica, Inc.
引用
收藏
页码:1360 / 1364
页数:5
相关论文
共 23 条
[1]   The influence of diabetes mellitus on acute and late clinical outcomes following coronary stent implantation [J].
Abizaid, A ;
Kornowski, R ;
Mintz, GS ;
Hong, MK ;
Abizaid, AS ;
Mehran, R ;
Pichard, AD ;
Kent, KM ;
Satler, LF ;
Wu, HS ;
Popma, JJ ;
Leon, MB .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1998, 32 (03) :584-589
[2]   Final results of a randomized trial comparing the NIR stent to the Palmaz-Schatz stent for narrowings in native coronary arteries [J].
Baim, DS ;
Cutlip, DE ;
O'Shaughnessy, CD ;
Hermiller, JB ;
Kereiakes, DJ ;
Giambartolomei, A ;
Katz, S ;
Lansky, AJ ;
Fitzpatrick, M ;
Popma, JJ ;
Ho, KKL ;
Leon, MB ;
Kuntz, RE .
AMERICAN JOURNAL OF CARDIOLOGY, 2001, 87 (02) :152-156
[3]   Paired comparison of vascular wall reactions to Palmaz stents, Strecker tantalum stents, and Wallstents in canine iliac and femoral arteries [J].
Barth, KH ;
Virmani, R ;
Froelich, J ;
Takeda, T ;
Lossef, SV ;
Newsome, J ;
Jones, R ;
Lindisch, D .
CIRCULATION, 1996, 93 (12) :2161-2169
[4]   Predictors of restenosis after coronary stent implantation [J].
Bauters, C ;
Hubert, E ;
Prat, A ;
Bougrimi, K ;
Van Belle, E ;
McFadden, EP ;
Amouyel, P ;
Lablanche, JM ;
Bertrand, M .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1998, 31 (06) :1291-1298
[5]  
Edelman ER, 2001, CIRCULATION, V103, P429
[6]   A RANDOMIZED COMPARISON OF CORONARY-STENT PLACEMENT AND BALLOON ANGIOPLASTY IN THE TREATMENT OF CORONARY-ARTERY DISEASE [J].
FISCHMAN, DL ;
LEON, MB ;
BAIM, DS ;
SCHATZ, RA ;
SAVAGE, MP ;
PENN, I ;
DETRE, K ;
VELTRI, L ;
RICCI, D ;
NOBUYOSHI, M ;
CLEMAN, M ;
HEUSER, R ;
ALMOND, D ;
TEIRSTEIN, PS ;
FISH, RD ;
COLOMBO, A ;
BRINKER, J ;
MOSES, J ;
SHAKNOVICH, A ;
HIRSHFELD, J ;
BAILEY, S ;
ELLIS, S ;
RAKE, R ;
GOLDBERG, S .
NEW ENGLAND JOURNAL OF MEDICINE, 1994, 331 (08) :496-501
[7]   Stent and artery geometry determine intimal thickening independent of arterial injury [J].
Garasic, JM ;
Edelman, ER ;
Squire, JC ;
Seifert, P ;
Williams, MS ;
Rogers, C .
CIRCULATION, 2000, 101 (07) :812-818
[8]   Patterns and mechanisms of in-stent restenosis - A serial intravascular ultrasound study [J].
Hoffmann, R ;
Mintz, GS ;
Dussaillant, GR ;
Popma, JJ ;
Pichard, AD ;
Satler, LF ;
Kent, KM ;
Griffin, J ;
Leon, MB .
CIRCULATION, 1996, 94 (06) :1247-1254
[9]   Intravascular ultrasound predictors of angiographic restenosis in lesions treated with Palmaz-Schatz stents [J].
Hoffmann, R ;
Mintz, GS ;
Mehran, R ;
Pichard, AD ;
Kent, KM ;
Satler, LF ;
Popma, JJ ;
Wu, HS ;
Leon, MB .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1998, 31 (01) :43-49
[10]   Angiographic and intravascular ultrasound predictors of in-stent restenosis [J].
Kasaoka, S ;
Tobis, JM ;
Akiyama, T ;
Reimers, B ;
Di Mario, C ;
Wong, ND ;
Colombo, A .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1998, 32 (06) :1630-1635