Label-Free Electrical Detection of Cardiac Biomarker with Complementary Metal-Oxide Semiconductor-Compatible Silicon Nanowire Sensor Arrays

被引:222
作者
Chua, Jay Huiyi [1 ]
Chee, Ru-Ern [1 ]
Agarwal, Ajay [1 ]
Wong, She Mein [1 ]
Zhang, Guo-Jun [1 ]
机构
[1] Agcy Sci Technol & Res, Inst Microelect, Singapore 117685, Singapore
关键词
TROPONIN-T; DNA HYBRIDIZATION; BIOSENSORS; NANOSENSORS; SPECIFICITY; MOLECULES; INJURY;
D O I
10.1021/ac901157x
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Arrays of highly ordered silicon nanowire (SiNW) clusters are fabricated using complementary metal-oxide semiconductor (CMOS) field effect transistor-compatible technology, and the ultrasensitive, label-free, electrical detection of cardiac biomarker in real time using the array sensor is presented. The successful detection of human cardiac troponin-T (cTnT) has been demonstrated in an assay buffer solution of concentration down to 1 fg/mL, as well as in an undiluted human serum environment of concentration as low as 30 fg/mL. The high specificity, selectivity, and swift response time of the SiNWs to the presence of ultralow concentrations of a target protein in a biological analyte solution, even in the presence of a high total protein concentration, paves the way for the development of a medical diagnostic system for point-of-care application that is able to provide an early and accurate indication of cardiac cellular necrosis.
引用
收藏
页码:6266 / 6271
页数:6
相关论文
共 28 条
[1]   CARDIAC TROPONIN-I - A MARKER WITH HIGH SPECIFICITY FOR CARDIAC INJURY [J].
ADAMS, JE ;
BODOR, GS ;
DAVILAROMAN, VG ;
DELMEZ, JA ;
APPLE, FS ;
LADENSON, JH ;
JAFFE, AS .
CIRCULATION, 1993, 88 (01) :101-106
[2]   Tissue specificity of cardiac troponin I, cardiac troponin T and creatine kinase-MB [J].
Apple, FS .
CLINICA CHIMICA ACTA, 1999, 284 (02) :151-159
[3]   Quantitative real-time measurements of DNA hybridization with alkylated nonoxidized silicon nanowires in electrolyte solution [J].
Bunimovich, Yuri L. ;
Shin, Young Shik ;
Yeo, Woon-Seok ;
Amori, Michael ;
Kwong, Gabriel ;
Heath, James R. .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 2006, 128 (50) :16323-16331
[4]   Nanowire nanosensors for highly sensitive and selective detection of biological and chemical species [J].
Cui, Y ;
Wei, QQ ;
Park, HK ;
Lieber, CM .
SCIENCE, 2001, 293 (5533) :1289-1292
[5]   Electrical conduction through DNA molecules [J].
Fink, HW ;
Schönenberger, C .
NATURE, 1999, 398 (6726) :407-410
[6]   Silicon nanowire arrays for label-free detection of DNA [J].
Gao, Zhiqiang ;
Agarwal, Ajay ;
Trigg, Alastair D. ;
Singh, Navab ;
Fang, Cheng ;
Tung, Chih-Hang ;
Fan, Yi ;
Buddharaju, Kavitha D. ;
Kong, Jinming .
ANALYTICAL CHEMISTRY, 2007, 79 (09) :3291-3297
[7]   Cardiac troponin T Isoforms affect the Ca2+ sensitivity and inhibition of force development -: Insights into the role of troponin T isoforms in the heart [J].
Gomes, AV ;
Guzman, G ;
Zhao, JJ ;
Potter, JD .
JOURNAL OF BIOLOGICAL CHEMISTRY, 2002, 277 (38) :35341-35349
[8]   Direct ultrasensitive electrical detection of DNA and DNA sequence variations using nanowire nanosensors [J].
Hahm, J ;
Lieber, CM .
NANO LETTERS, 2004, 4 (01) :51-54
[9]   THE PROGNOSTIC VALUE OF SERUM TROPONIN-T IN UNSTABLE ANGINA [J].
HAMM, CW ;
RAVKILDE, J ;
GERHARDT, W ;
JORGENSEN, P ;
PEHEIM, E ;
LJUNGDAHL, L ;
GOLDMANN, B ;
KATUS, HA .
NEW ENGLAND JOURNAL OF MEDICINE, 1992, 327 (03) :146-150
[10]   Ultrasensitive, label-free, and real-time immunodetection using silicon field-effect transistors [J].
Kim, Ansoon ;
Ah, Chil Seong ;
Yu, Han Young ;
Yang, Jong-Heon ;
Baek, In-Bok ;
Ahn, Chang-Guen ;
Park, Chan Woo ;
Jun, Myung Sim ;
Lee, Sungjae .
APPLIED PHYSICS LETTERS, 2007, 91 (10)